![THORLABS BC207UV Operating Manual Download Page 159](http://html.mh-extra.com/html/thorlabs/bc207uv/bc207uv_operating-manual_1110819159.webp)
BC207 Series
Chapter 8: Maintenance and Repair
Rev: 8.1, 10-Jan-2022
Page 152
© 2022 Thorlabs
8.3.1 Warnings and Errors
Warnings and error messages will appear in the status bar as soon as improper measurement
conditions are detected. For this reason, please keep always an eye on the status bar in order
to prevent measurement errors. Below some examples are shown:
q
Power too high, camera saturated!
Explanation:
Some pixels within the selected ROI or Calculation Area are saturated because
they reach the maximum digit value 4095 in Precision Mode (12 bit data))..
Therefore, the local beam intensity may be even higher than displayed. Beam
parameters cannot be calculated correctly!
Resolving:
Decrease Exposure Time and Gain settings or reduce the beam intensity using
a higher attenuation ND filter of the
. You may also cascade two
filters. If Exposure Control is set to Auto, a short appearance of this error is
harmless as Exposure Time and Gain will be adjusted automatically.
5