© 2015 Thorlabs
84
Series 4000 SCPI Programmers Reference Manual
4.3 Device Specific Errors
Error
Description
-310
System error
Indicates that some error, termed “system error” by the device, has occurred.
-311
Memory error
Indicates some physical fault in the device’s memory, such as parity error.
-313
Calibration memory lost
Indicates that nonvolatile calibration data used by the
*CAL?
command has been lost.
-314
Save/recall memory lost
Indicates that the nonvolatile data saved by the
*SAV?
command has been lost.
-315
Configuration memory lost
Indicates that nonvolatile configuration data saved by the device has been lost.
-321
Out of memory
An internal operation needed more memory than was available.
-330
Self-test failed
There was an error detected during a self-test. During a self-test the instrument checks the internal
hardware for proper function.
-340
Calibration failed
The instrument is not or improper calibrated.
-350
Queue overflow
A specific code entered into the queue in lieu of the code that caused the error. This code indicates
that there is no room in the queue and an error occurred but was not recorded.
-363
Input buffer overrun
Software or hardware USB input port buffer overflows with data caused by improper or nonexistent
pacing.
4.4 Query Errors
Error
Description
-400
Query error
This code indicates that a generic Query Error as defined in IEEE 488.2, 11.5.1.1.7 and 6.3 has
occurred.
-410
Query INTERRUPTED
Indicates that a condition causing an INTERRUPTED Query error occurred (see IEEE 488.2, 6.3.2.3);
for example, a query followed by another query before a response was completely sent.
Summary of Contents for 4000 Series
Page 1: ...Series 4000 LDC40xx TED4015 ITC40xx SCPI Programmer s Reference Manual 2015 ...
Page 2: ...Version Date 3 3 08 Apr 2015 Copyright 2015 Thorlabs ...
Page 69: ... 2015 Thorlabs 3 Commands by Subsystem 67 Figure FIGURENO Status Structure ...
Page 92: ... 2015 Thorlabs 90 Series 4000 SCPI Programmers Reference Manual instrument ...