© 2015 Thorlabs
30
Series 4000 SCPI Programmers Reference Manual
3.4.3 Polarity
Syntax
OUTPut[1]:POLarity {NORMal|CG|INVerted|AG}
OUTPut[1]:POLarity?
Description
The
OUTPut[1]:POLarity
command sets the LD output polarity to cathode ground (
NORMal
or
CG
) or anode ground (
INVerted
or
AG
). The parameters
NORMal
and
INVerted
are for SCPI
conformance.
Example
OUTP:POL CG
Sets the LD output polarity to cathode ground.
OUTP:POL?
Queries the LD output polarity.
CG
The response is either
CG
or
AG
.
Note
This command is valid for LDC4000 and ITC4000 Series instruments only.
3.4.4 Noise Reduction Filter
Syntax
:FILTer[:LPASs]:STATe {ON|1|OFF|0}
:FILTer[:LPASs]:STATe?
Description
The
FILTer[:LPASs]:STATe
command switches the LD output low pass filter on or off.
Example
FILT:STAT ON
Switches LD output noise reduction filter ON.
FILT:STAT?
Queries the LD output noise filter status.
The response is ASCII
1
if the filter is switched on or an ASCII
0
if the output is off.
Note
This command is valid for LDC4005, ITC4001, ITC4002QCL, ITC4005 and ITC4005QCL
instruments, FW version 1.5 up, only.