4
System Monitor
4-26
HeartMate III Left Ventricular Assist System Instructions for Use
Hazard Alarms
IMPORTANT!
If communication is lost between the System Controller and the Pump, this causes
a Controller Fault Hardware hazard alarm. The System Monitor will not display that alarm. Only
the System Controller displays the Controller Fault Hardware alarm.
There are five hazard alarms that display on the System Monitor. These hazard alarms are listed
in order of priority:
• PUMP OFF: The Pump has been turned off or its Driveline is disconnected from the
System Controller. Immediately turn on the Pump or reconnect the Driveline to the
System Controller.
• DRIVELINE DISCONNECTED: The Driveline is disconnected from the System Controller.
Immediately reconnect the Driveline to the System Controller.
• NO EXTERNAL POWER: The System Controller is not receiving external power from either
power cable, and Pump function is being supported by the System Controller’s backup battery.
Immediately connect the System Controller to a power source (Power Module, Mobile Power
Unit, or two HeartMate 14 Volt Lithium-Ion batteries).
• LOW FLOW: Pump flow is less than 2.5 lpm, the Pump has stopped, the Pump is not operating
properly, or the Driveline is disconnected from the System Controller. The count up time listed in
the parameters box indicates how long this alarm has been active. Changes in patient
conditions can result in low flow, such as hypertension.
• LOW VOLTAGE: Less than five minutes of combined power remain for the two HeartMate
14 Volt Lithium-Ion batteries (during battery-powered operation), or the System Controller is
receiving inadequate power from the Power Module or Mobile Power Unit. Immediately
connect the System Controller to a functioning power source.
Summary of Contents for HeartMate III
Page 1: ...Instructions for Use ...
Page 2: ......
Page 3: ...Thoratec Corporation HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM Instructions for Use ...
Page 4: ...HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE ii ...
Page 6: ...HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE iv ...
Page 10: ...Contents viii HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE ...
Page 12: ...1 Introduction 1 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 28: ...1 Introduction 1 18 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 30: ...2 System Operations 2 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 88: ...3 Powering the System 3 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 112: ...3 Powering the System 3 26 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 166: ...4 System Monitor 4 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 208: ...4 System Monitor 4 44 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 210: ...5 Surgical Procedures 5 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 378: ...A A 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 390: ...B B 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 420: ...C C 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 426: ...C C 8 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 428: ...D D 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 448: ...E E 10 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 457: ......