4
System Monitor
4-20
HeartMate III Left Ventricular Assist System Instructions for Use
6. Determine the minimum Fixed Speed:
a. Starting from the current Fixed Speed, lower the speed gradually to a value as low as
possible without the patient experiencing signs of worsening heart failure.
For example, shortness of breath or lightheadedness.
b. Allow the patient to stabilize at each speed setting.
IMPORTANT!
Do not allow the Fixed Speed to drop below 3,000 rpm under any
circumstances.
c.
Reduce the speed until the aortic valve opens with each beat or the patient starts to
become symptomatic.
d. Record the patient’s current heart rate, blood pressure, and Pump speed.
e.
Using echocardiography, record the patient’s left ventricular diameter and position of the
septum.
7.
Determine the maximum Fixed Speed:
a. Starting from the minimum Fixed Speed (determined in Step 6), increase the pump speed
gradually until echocardiography shows a flattening of the interventricular septum (or is
clinically acceptable based on the echocardiographic evaluation).
b. Record the patient’s current heart rate, blood pressure, and Pump speed.
c.
Using echocardiography, record the patient’s left ventricular diameter and frequency of
aortic valve opening.
8. Based on findings from the speed study, determine the optimum Fixed Speed, which usually
falls midway between the minimum and maximum speeds.
IMPORTANT!
The selected speed may be adjusted based on clinical judgment regarding the
need for periodic aortic valve opening and a palpable pulse. To accommodate normal shifts in
volume and hemodynamic status, the Fixed Speed should generally be set at least 400 rpm below
the maximum Fixed Speed determined previously.
Low Speed Limit
The Low Speed Limit should be set at the lowest speed at which the Pump can run while
maintaining patient stability. It is important to establish a Low Speed Limit that can sustain a
patient safely. Establishing a Low Speed Limit will maximize the protective benefits during PI
events or when the Pump is in Power Saver mode.
In Power Saver mode, the System Controller slows the Pump speed to save power. If power is
removed or fails, the System Controller provides 15 minutes of full power before entering
Power Saver mode. Note that alarms cannot be silenced while the System Controller is in
Power Saver mode.
Summary of Contents for HeartMate III
Page 1: ...Instructions for Use ...
Page 2: ......
Page 3: ...Thoratec Corporation HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM Instructions for Use ...
Page 4: ...HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE ii ...
Page 6: ...HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE iv ...
Page 10: ...Contents viii HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE ...
Page 12: ...1 Introduction 1 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 28: ...1 Introduction 1 18 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 30: ...2 System Operations 2 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 88: ...3 Powering the System 3 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 112: ...3 Powering the System 3 26 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 166: ...4 System Monitor 4 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 208: ...4 System Monitor 4 44 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 210: ...5 Surgical Procedures 5 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 378: ...A A 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 390: ...B B 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 420: ...C C 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 426: ...C C 8 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 428: ...D D 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 448: ...E E 10 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 457: ......