Powering the System
3
HeartMate III Left Ventricular Assist System Instructions for Use
3-37
Mobile Power Unit User Interface Components
describes the buttons, lights, symbols, and display screen on the Mobile Power Unit
user interface.
Setting Up the Mobile Power Unit
To set up the Mobile Power Unit, perform these tasks:
• Install the Mobile Power Unit batteries
• Connect the Mobile Power Unit power cord to the Mobile Power Unit and AC power
Installing or Replacing the Mobile Power Unit Batteries
The Mobile Power Unit uses three Alkaline AA batteries to power its alarms. The Mobile Power
Unit batteries must be installed before using the Mobile Power Unit. The batteries power the
alarm echo function when an AC power failure occurs or the power cord is disconnected.
The yellow Mobile Power Unit battery symbol (
) illuminates and a beeping audio tone sounds
when the Alkaline AA batteries are not installed, or when the batteries are depleted and need to
be replaced.
Table 3.3 Mobile Power Unit User Interface Components
Power On
Symbol
The power symbol is illuminated green when the Mobile Power Unit is powered
and functioning properly.
Yellow Wrench
Alarm
The yellow wrench symbol illuminates when the Mobile Power Unit detects a
mechanical, electrical, or software issue with the system.
This is an
Advisory
alarm. When the yellow wrench illuminates, switch to two
fully-charged HeartMate 14 Volt Lithium-Ion batteries.
Replace Mobile
Power Unit
Battery Alarm
The yellow Replace Mobile Power Unit Battery symbol illuminates when the
Alkaline AA batteries are not installed, or are depleted and need to be
replaced.
This is an
Advisory
alarm. When the Replace Mobile Power Unit Battery
symbol illuminates, replace the internal batteries in the Mobile Power Unit.
CAUTION !
Never change the Mobile Power Unit batteries while the Mobile Power Unit is powering
the HeartMate system. Switch to another power source, and then disconnect the Mobile
Power Unit power cord from the power socket prior to replacing the batteries.
Summary of Contents for HeartMate III
Page 1: ...Instructions for Use ...
Page 2: ......
Page 3: ...Thoratec Corporation HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM Instructions for Use ...
Page 4: ...HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE ii ...
Page 6: ...HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE iv ...
Page 10: ...Contents viii HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE ...
Page 12: ...1 Introduction 1 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 28: ...1 Introduction 1 18 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 30: ...2 System Operations 2 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 88: ...3 Powering the System 3 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 112: ...3 Powering the System 3 26 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 166: ...4 System Monitor 4 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 208: ...4 System Monitor 4 44 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 210: ...5 Surgical Procedures 5 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 378: ...A A 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 390: ...B B 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 420: ...C C 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 426: ...C C 8 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 428: ...D D 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 448: ...E E 10 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 457: ......