3
Powering the System
3-28
HeartMate III Left Ventricular Assist System Instructions for Use
Storing and Shipping the Power Module
If the Power Module is without electrical power for approximately 18 hours or more, the internal
backup battery may be damaged. The Power Module backup battery must be disconnected if the
Power Module is not being used and will be unplugged from electrical power for an extended
time (such as for travel or for transport for service or maintenance). Disconnecting the Power
Module backup battery prevents damage to the battery.
Disconnecting the Power Module Backup Battery
The Power Module backup battery should be disconnected any time the Power Module is
unplugged for an extended period, such as when the Power Module is shipped for service or
during extended periods of time without AC. In these situations, the battery remains in the battery
compartment but is not connected.
F
OR
THIS
TASK
YOU
NEED
:
• 1 Power Module backup battery installed and connected
• 1 crosshead (Phillips) screwdriver
T
O
D
ISCONNECT
THE
P
OWER
M
ODULE
BACKUP
BATTERY
:
1. Transfer the patient to battery power prior to disconnecting the Power Module backup battery.
2. Place the Power Module on a flat, stable surface.
3. Ensure that the Power Module is unplugged from AC power and disconnected from the patient.
4. Inspect the Power Module for dents, chips, cracks, or other signs of damage, and complete
one of the following steps:
• If the Power Module is undamaged, go to Step 5.
• If the Power Module is damaged, contact Thoratec Corporation for a replacement if
needed. Do not use a Power Module that appears damaged.
5. Use a crosshead (Phillips) screwdriver to loosen the two ¼-turn screws from the rear panel.
The screws remain in the screw holes to ensure they are not lost (
Figure 3.31 Loosening the Screws
Summary of Contents for HeartMate III
Page 1: ...Instructions for Use ...
Page 2: ......
Page 3: ...Thoratec Corporation HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM Instructions for Use ...
Page 4: ...HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE ii ...
Page 6: ...HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE iv ...
Page 10: ...Contents viii HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE ...
Page 12: ...1 Introduction 1 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 28: ...1 Introduction 1 18 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 30: ...2 System Operations 2 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 88: ...3 Powering the System 3 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 112: ...3 Powering the System 3 26 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 166: ...4 System Monitor 4 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 208: ...4 System Monitor 4 44 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 210: ...5 Surgical Procedures 5 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 378: ...A A 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 390: ...B B 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 420: ...C C 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 426: ...C C 8 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 428: ...D D 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 448: ...E E 10 HeartMate III Left Ventricular Assist System Instructions for Use ...
Page 457: ......