DFT-6600 Battery Charger-Discharger Users Manual
22
a)
The Stop voltage set in the parameter setting plays a role in protecting the battery, which’s correct setting of
the stop voltage value can avoid overcharge or over-discharge of the battery and avoid damage to the
battery.
Follows set value for reference:
2V battery - discharge cutoff voltage: 1.80V, charging cutoff voltage: 2.20V;
6V battery - discharge stop voltage: 5.40V, charging stop voltage: 6.60V;
12V battery - discharge stop voltage: 10.8V, charging stop voltage: 13.20V;
b)
For small capacity batteries, you should also pay attention to the setting of the working current. It is
recommended to charge/discharge at a rate of 0.1C (i.e. the operating current is 0.1 times of the battery
capacity value).
4.4. Discharge & Charge Cycle (optional)
You can also set the Discharge & Charge Cycle Nose for Automatically Discharging and Charging together in one.
4.5. Data management
Test results in the data management menu. Can be exported and delete.
Data management includes details discharge measurement data & charging measurement data.
4.5.1 Data viewing
Press "Data Management" in the main menu, and select the data type, select the data, then press the [Open] key,