11.2.23
Command TR
<id>TR<parameter><CR>
Measured value request.
Access:
reading
Description:
The command activates the one-time transfer of the current
data telegram.
Parameter type:
unsigned integer
Parameter:
1:
Query measured value telegram 1
2:
Query measured value telegram 2
4:
Query measured value telegram 4
5:
Query measured value telegram 5
6:
Query measured value telegram 6
30:
Query measured value memory telegram
Type return value:
Character string
Return value:
Value range:
1 / 2 / 30
Initial value:
-
11.2.24
Command UC
<id>UC<parameter><CR> Selection calibration table.
Access:
reading / writing
Description:
The command selects between the standard curve and the
(most recent) calibration table with the highest value. Parameter
UC can be set to one only when a calibration table is available.
Parameter type:
unsigned integer
Parameter:
0: Standard curve
1: (most recent) calibration table with the highest value
Type return value:
unsigned integer
Return value:
see parameter
Value range:
0 / 1
Initial value:
0