IMPORTANT NOTICE
Temperature measurements made using the FevIR Scan 2 system should not be solely
or primarily relied upon to diagnose or exclude a diagnosis of COVID-19 or any other
disease.
An elevated body temperature in the context of use must be confirmed by a suitably
trained person with secondary evaluation methods (e.g., an NCIT or clinical grade
contact thermometer)
Operation of the system must only be carried out with the ThermaRef 35B1
blackbody supplied in operation and placed at the same point (same horizontal
plane) and distance from the camera as the subject inspection point to comply with
system accuracy and performance. The inspection point must be between 1 and 3m
(3.2 to 9.5ft) from the camera. The system self-check and continuous self-test will
show a “System Not Ready” on screen message and a warning triangle which will
prevent system operation if the blackbody is not present, operational, or withing the
field of view.
Only one person at a time must be presented to the camera at the inspection point
The FevIR Scan 2 system has an automatically operated, self-calibrating, Non-
Uniformity Correction (NUC) calibration shutter fitted in front of the thermal lens,
immediately inside the front face of the camera. Do not insert any object into this
opening or attempt to interfere with the operation of the shutter. Any interference
to the operation of the shutter may cause permanent damage to the unit and render
the Manufacturer’s Warranty invalid.
The FevIR Scan 2 camera model FSD01 and the ThermaRef 35B1 units must be re-
calibrated annually from the last date of calibration. This date is shown on the
certificates supplied with the equipment and by a label fixed to each device. Please
contact Thermoteknix for assistance.