In the example above the oven temperature displayed (PV) is 100ºC. If a certified reference 
instrument measured 99ºC we would need to correct the temperature by -1.0ºC (note that the 
sensor correction parameter is accurate to 1 decimal place). 

 
 
MAINTENANCE 
 

The Thermoline Scientific Laboratory Ovens require no routine maintenance other than normal 
levels of cleanliness. 
 
The external surface powder coated steel and the inner surface and door stainless steel may 
be wiped clean using a damp, soft cloth. 
 

CARING FOR STAINLESS STEEL 

Stainless steel is under most conditions extremely resistant to corrosion. This is in part due to 
the addition of chromium and nickel to the steel and the formation of a durable chromium oxide 
at the surface during the manufacturing process. There are several chemicals which will attack 
the surface of stainless steel, plus the lack of oxygen at the surface will cause rusting, 
corrosion and pitting. 

Note: 

Deionised water is particularly corrosive, to minimise, maintain a 

Neutral pH (7-9).  
Always have in place a program of regular cleaning using a soft damp cloth with a mild solution 
of soap and water and allow to dry. 
 

TECHNICAL ASSISTANCE 
 

If you require additional operational or technical information regarding this equipment please contact: 
 
Thermoline Scientific 
Customer Service Division 
Telephone: 

 61 2 9604 3911 

Facsimile:  

 61 2 9725 1706 

Email: 

[email protected]

 

 

 

Press the page 

button to access 

the sensor 

correction 

parameter. 

Use the up or 

down arrow to 

enter the sensor 

correction value. 

Allow the digit to 

stop flashing and 

the screen will 

display the adjusted 

value. 

Summary of Contents for TO-152G

Page 1: ...ox 6862 Wetherill Park Delivery Centre N S W 1851 Australia Phone 02 9604 3911 International 61 2 9604 3911 Fax 02 9725 1706 International 61 2 9725 1706 Email sales thermoline com au Web www thermoline com au LABORATORY OVENS FAN FORCED GRAVITY CONVECTION MODELS TO 152G TOH 152G TO 152F TOH 152F TO 235F TO 330F TO 500F WITH OMRON E5CC CONTROL ...

Page 2: ...e packing material and check for damage Notify the detail of any damage to your supplier or to Thermoline Scientific without delay Retain the packing materials until the equipment has been thoroughly tested Inventory Shelves 150 litre models 3 shelves 235 330 litre models 4 shelves 500 litre model 5 shelves Operating Instructions Electrical mains cable USE FUNCTION The Thermoline range of Premium ...

Page 3: ...ying purposes but should not be used with large volumes of water since the interior is not sealed A range of dehydrating ovens with sealed liners is available for large capacity drying purposes ELECTRICAL These ovens are suitable for connection to a standard 240 volt 50Hz supply A dedicated outlet should be used for the supply do not use power boards or the like A 3 pin moulded plug is fitted as s...

Page 4: ...rsor when changing temperature Down arrow Used to decrease parameter setting Up arrow Used to increase parameter setting OPERATION 1 Locate the oven as previously described in Location plug the mains lead into the power supply and turn on the power point 2 Turn on the main switch on the control panel The switch is illuminated so you will know when it is on The temperature will go through a short w...

Page 5: ...ccuracy of the temperature displayed in relation to the actual temperature inside the oven these include the following Product load inside the oven the load should be distributed evenly across each shelf with room for vertical air flow around the product Oven temperature at higher temperatures the heat loss from the oven will be greater Location of the sensor the temperature sensor can never be pl...

Page 6: ...a durable chromium oxide at the surface during the manufacturing process There are several chemicals which will attack the surface of stainless steel plus the lack of oxygen at the surface will cause rusting corrosion and pitting Note Deionised water is particularly corrosive to minimise maintain a Neutral pH 7 9 Always have in place a program of regular cleaning using a soft damp cloth with a mil...

Page 7: ...al number This is generally found on the exterior of the cabinet in the form of a stick on label The company name address contact name contact phone number A brief description of the problem All warranty claims must be reported to and agreed to by a Thermoline representative prior to any work being carried out MODEL NO SERIAL NO ...

Page 8: ...tanding the preceding clause and to the extent permissible by law the liability of Thermoline is limited in relation to the warranted product and at the option of Thermoline to replacing the product or the supply of equivalent product the repair of the product the payment of the cost of replacing the product or of acquiring equivalent product or the payment of the cost of having the product repair...

Reviews: