background image

P a g e

 | 

 

26-1-16 

On the fan forced models a circulating fan blows the air across a heating element mounted 
behind the ceiling panel, down a rear air duct, and across each shelf level for even 
temperature distribution. 
On gravity convection models the element is located at the bottom of the oven. 
The temperature sensor provides an accurate measure of the temperature inside the oven. 

 
 
 
 

CONTROL PANEL: 

 

 

Page button:

 Used to view calibration offset parameter. 

 

 

Scroll button:

 Not used by the operator. 

 

 

Side arrow:

 Used to move the cursor when changing temperature. 

 

    

Down arrow:

 Used to decrease parameter setting. 

 

    

Up arrow:

 Used to increase parameter setting. 

 

OPERATION: 

1.  Locate the oven as previously described in “Location”, plug the mains lead into the 

power supply and turn on the power point. 
 

2.  Turn on the main switch on the control panel. The switch is illuminated so you will know 

when it is on. The temperature will go through a short warm up period where all 
segments of the display will be on, before indicating the set temperature (SV) on the 
lower display and oven temperature (PV) on the top display. 
 

3.  The Omron E5CC is a microprocessor based instrument that has been factory 

configured for range, sensor type, and engineering parameters for optimum control. 

Summary of Contents for TO-152G

Page 1: ...ox 6862 Wetherill Park Delivery Centre N S W 1851 Australia Phone 02 9604 3911 International 61 2 9604 3911 Fax 02 9725 1706 International 61 2 9725 1706 Email sales thermoline com au Web www thermoline com au LABORATORY OVENS FAN FORCED GRAVITY CONVECTION MODELS TO 152G TOH 152G TO 152F TOH 152F TO 235F TO 330F TO 500F WITH OMRON E5CC CONTROL ...

Page 2: ...e packing material and check for damage Notify the detail of any damage to your supplier or to Thermoline Scientific without delay Retain the packing materials until the equipment has been thoroughly tested Inventory Shelves 150 litre models 3 shelves 235 330 litre models 4 shelves 500 litre model 5 shelves Operating Instructions Electrical mains cable USE FUNCTION The Thermoline range of Premium ...

Page 3: ...ying purposes but should not be used with large volumes of water since the interior is not sealed A range of dehydrating ovens with sealed liners is available for large capacity drying purposes ELECTRICAL These ovens are suitable for connection to a standard 240 volt 50Hz supply A dedicated outlet should be used for the supply do not use power boards or the like A 3 pin moulded plug is fitted as s...

Page 4: ...rsor when changing temperature Down arrow Used to decrease parameter setting Up arrow Used to increase parameter setting OPERATION 1 Locate the oven as previously described in Location plug the mains lead into the power supply and turn on the power point 2 Turn on the main switch on the control panel The switch is illuminated so you will know when it is on The temperature will go through a short w...

Page 5: ...ccuracy of the temperature displayed in relation to the actual temperature inside the oven these include the following Product load inside the oven the load should be distributed evenly across each shelf with room for vertical air flow around the product Oven temperature at higher temperatures the heat loss from the oven will be greater Location of the sensor the temperature sensor can never be pl...

Page 6: ...a durable chromium oxide at the surface during the manufacturing process There are several chemicals which will attack the surface of stainless steel plus the lack of oxygen at the surface will cause rusting corrosion and pitting Note Deionised water is particularly corrosive to minimise maintain a Neutral pH 7 9 Always have in place a program of regular cleaning using a soft damp cloth with a mil...

Page 7: ...al number This is generally found on the exterior of the cabinet in the form of a stick on label The company name address contact name contact phone number A brief description of the problem All warranty claims must be reported to and agreed to by a Thermoline representative prior to any work being carried out MODEL NO SERIAL NO ...

Page 8: ...tanding the preceding clause and to the extent permissible by law the liability of Thermoline is limited in relation to the warranted product and at the option of Thermoline to replacing the product or the supply of equivalent product the repair of the product the payment of the cost of replacing the product or of acquiring equivalent product or the payment of the cost of having the product repair...

Reviews: