Section 6 - Specifications
6.1 Model 8018 - Controller Specifications
Dimensions
14”W x 6”H* x 9” F-B
(35.6cm x 15.2cm x 22.9cm)
*Add 7” (18cm) for stand
Weight
20 lbs (9 kg)
Electrical
95-125VAC, 50/60Hz, 2 FLA
(190-250VAC, 50/60Hz)
Temperature range
+50°C to -180°C (+122°F to -
290°F)
Rates
.01°C per minute to 99.9°C/minute
Hold times
1 minute to 999 minutes
Error Limits
Time
1 second to 999 minutes
Temperature
1°C to 180°C (34°F to -290°F)
6.2 Chamber Specifications
8022 (8023)
8024 (8025)
8026 (8027)
External dimensions
10”W x 14.25H x 15.25 F-B
(25.4cm x 36.2cm x 38.7cm)
17.75”W x 16.365H x 16.5” F-B
(45cm x 40.7cm x 41.9cm)
24.5”W x 18”H x 13” F-B
(62.2cm x 45.7cm x 50.8cm)
Internal dimensions
6” x 11.38H x 5” F-B
(15.2cm x 28.9cm x 12.7cm)
9”W x 12”H x 9” F-B
(22.9cm x 30.5cm x 22.9cm)
13”W x 13”H x 13” F-B
(33cm x 33cm x 33cm)
Volume
353 cu. in.
972 cu. in.
2197 cu. in.
Weight 30 lbs (13 kg)
51 lbs (23 kg)
90 lbs (41 kg)
Electrical (supplied by controller)
95-125VAC, 50/60 Hz
Same
Same
(190-250VAC, 50/60 Hz)
8022 (8023)
8024 (8025) 8026 (8027)
Fan
Diameter 4” (10.2cm)
8” (20cm)
8” (20.3cm)
# of wings
4
5 5
RPM’s
2880 1550 1550
Sensing element
Mounted chamber thermocouple Type T
Flexible
Sample thermocouple Type T
Utility connections
1/2” (1.3cm) 45° Flare
4 ft. (122cm) or 6 ft. (183cm)
flexible hose with 1/2” (1.3cm)
flared tube fittings.*
*Connects directly to low pressure (22PSI) 180/230 liter LN2
supply tank, or available 35 and 50 liter supply tanks.
Models 1010 and 1011_________________________________________________________________________Specifications
6 - 1