Important Operating Parameters
The parameters described in this section should be considered for
routine operation of the detector. You can usually access these
parameters from the Chromeleon user interface. If a parameter listed
below is not available in Chromeleon, consider updating the firmware
and Chromeleon version.
TIP
The Instrument Method Wizard provides different parameter view
modes for the detector. Depending on the parameters to be set, you can
select the desired view mode (for example,
Easy
or
Advanced
).
For more information, refer to
Chromeleon Help and documents
.
Parameters
Description
UV lamp
Turn on the lamp before starting an analysis
(
UV_Lamp = On
). This setting allows turning on and off the
lamp of the detector.
Shutter
Move the shutter (filter paddle) into the required position:
•
Open position for data acquisition
Check that flow is established through the flow cell
before you open the shutter while the lamp is turned on.
•
Closed position for protection of the flow cell when no
flow is established through the flow cell and the lamp is
turned on.
UV_VIS
The
UV_VIS
signal channels (
UV_VIS_x
) that are available
for data acquisition are listed in the dialog box for the
detector in the Instrument Configuration Manager
(Chromeleon 7) or Server Configuration program
(Chromeleon 6.8). Select the required signal channels.
The detector can record up to 10 signal channels
simultaneously, thus at 10 different wavelengths. Most
measurement parameters, such as the wavelength, are
settable for each signal channel separately.
Acquisition On
Acquisition Off
Turn on or off the data acquisition.
TIP
When you start the data acquisition or perform an
autozero, the shutter in the light path before the flow cell
is opened automatically. After the data acquisition, close
the shutter, if required.
Summary of Contents for Vanquish VH-D10
Page 4: ...Contacting Us Page 4 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 44: ...4 Unpacking Page 44 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 80: ...5 Installation Page 80 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 110: ...6 Operation Page 110 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 162: ...7 Maintenance and Service Page 162 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 190: ...11 Appendix Page 190 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 191: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 191 12 Index...
Page 195: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 195...