4. Check the waste line over the entire flow path:
a) Make sure that no bending (kink), pinching or squeezing of the
waste line is present at any point in the flow path.
b) If you have to cut tubing to length, use a tubing cutter. Make
sure that the cut is at right angle to the length of the line.
When cutting the waste line to length, also make sure that the
cut is not crimped.
The waste line should go straight to the system base and to
waste. Make sure that the line is positioned straight in the
tubing guides.
Figure 22: Flow connections on the LightPipe flow cell
5.6.5
Guiding Liquid Leaks to Waste
Leaking liquids of the detector are collected in the leak tray, where they
flow off through the chase on the right side of the leak tray to the drain
system.
For information about how the liquid is discharged to waste through the
Vanquish drain system, refer to the
Vanquish System Operating Manual
.
Summary of Contents for Vanquish VH-D10
Page 4: ...Contacting Us Page 4 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 44: ...4 Unpacking Page 44 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 80: ...5 Installation Page 80 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 110: ...6 Operation Page 110 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 162: ...7 Maintenance and Service Page 162 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 190: ...11 Appendix Page 190 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 191: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 191 12 Index...
Page 195: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 195...