Troubleshooting
Fluorescence Detectors (VC-D50, VC-D51, VF-D50, VF-D51)
Operating Manual
Page 163
4. Allow the sensor to adjust to the ambient temperature for a few
minutes.
5. If leakage is no longer reported, you can resume operation.
See also
2
2
8.3.2
Additional Device Operating Issues
This section provides additional issues that may arise during operation of
the Vanquish device. Locate the table for the type of symptom you have,
find the possible cause, and use the description of the solution to help
you solve your problem quickly.
Also check the Chromeleon Audit Trail for a related message if an
operating problem occurs. The message may provide additional
information.
Note that this section provides information on symptoms and causes
directly related to the Vanquish device. For information about
troubleshooting for the Vanquish system, refer to the
Vanquish System
Operating Manual
.
TIP
If you are unable to resolve a problem following the instructions
given here or if you experience problems that are not covered in this
section, contact Thermo Fisher Scientific Technical Support for
assistance.
Symptom
Possible Cause
Remedial Action
Peak tailing
Fluorescence detector
flow cell inlet and outlet
exchanged
Verify that the flow cell inlet and outlet in a
fluorescence detector are connected correctly.
Negative peaks
Absorption/fluorescence
of analyte lower than of
mobile phase
Select different UV or fluorescence detection
wavelengths.
Use a mobile phase with less background
absorption/ fluorescence.
Dissolve the sample in mobile phase.
Fluorescence of the
substance or eluent is
quenched by other
components
Consider using the negative peaks for
quantification.
Markers in fluorescence
signal
The Sensitivity was
automatically reduced
Select a lower Sensitivity for your application. The
sample audit trail informs to which level the
Sensitivity was reduced to.
Summary of Contents for Vanquish VC-D50
Page 4: ...Contacting Us Page 4 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 10: ...Contents Page 10 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 30: ...2 Safety Page 30 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 42: ...3 Device Overview Page 42 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 88: ...5 Installation Page 88 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 122: ...6 Operation Page 122 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 170: ...9 Specifications Page 170 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 187: ...Index Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual Page 187...