6
•
Operation
Page 118
Fluorescence Detectors (VC-D50, VC-D51, VF-D50, VF-D51)
Operating Manual
6.10 Shutting Down the Device
If the device will not be operated for some time, follow the instructions
in this section to shut down the device.
TIP
The Chromeleon software provides procedures for automatically
preparing the chromatography system for shutdown. The procedures
include, for example, operations for reducing the flow rate, reducing the
temperature in temperature-controlled devices, and turning off the
detector lamps. For information about
Smart Shutdown
and
Smart
Standby
, refer to the
Chromeleon Help
.
6.10.1 Short-Term Shutdown (Interruption of Operation)
To interrupt operation of the device for a short period (short-term
shutdown), for example, overnight, observe these guidelines for the
Vanquish system modules, as required by your system arrangement:
•
For your Vanquish detector, note the following:
Detector Type
Description
Charged aerosol
detector
Check that sufficient gas is available to continue gas flowing
through the detector. This is to prevent any build-up of
residue from solvents or analytes. Gas must be flowing when
pump flow is delivered to the detector.
UV/VIS detectors
The lamp(s) in the detector can remain turned on.
Variable wavelength detector and VH-D10 diode array
detector only:
The shutter can be moved to a closed position
for protection of the flow cell.
Fluorescence
detector
Turn off temperature control for the flow cell.
•
Apply a flow of 0.05 mL/min and have the pump deliver an
appropriate solvent.
Check the lower pressure limit for the pump and adapt the value if
necessary. If the pressure falls below the lower limit, the pump stops
the flow.
•
Set the injection valve in the autosampler to the Inject position.
•
Make sure that the temperature of the column does not exceed
40 °C.
•
When resuming operation, let the flow equilibrate and verify that
the operating parameters for the other system modules are set as
required before proceeding.
Summary of Contents for Vanquish VC-D50
Page 4: ...Contacting Us Page 4 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 10: ...Contents Page 10 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 30: ...2 Safety Page 30 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 42: ...3 Device Overview Page 42 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 88: ...5 Installation Page 88 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 122: ...6 Operation Page 122 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 170: ...9 Specifications Page 170 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 187: ...Index Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual Page 187...