Thermo Scientific
imPULSE S.U.M. with Touchscreen Console User’s Guide
|
66
Chapter 3
|
Touchscreen Console setup
3.4.9 Configuring the Harvest module
The Mass module must be configured and pumps must be set up prior
to configuring the Harvest module. Use the following information and
steps to configure the Harvest module.
To begin, locate the Harvest module on the Touchscreen Console
Home screen, and touch
Configure
. The Harvest module Configuration
screen will appear (Figure 3.39).
Note:
You can touch
Done
at the bottom of the Configuration screen
at any time to save your current settings and so you can resume later.
4. Enter your desired values for the following fields:
•
Fluid target:
The total amount of liquid you would like to pump into
the BPC.
Note:
This is the only option available if there is not at
least one pump set up.
•
Stage 1 flow rate:
The rate you would like to use when initially
filling the BPC (in L or kg per minute).
•
% Filled before stage 2:
The percentage of liquid that should be
pumped into the BPC before the pump moves to Stage 2.
•
Stage 2 flow rate:
The rate you would like to use when finishing
the filling process. The Stage 2 flow rate should be significantly
slower than Stage 1 to prevent overfilling.
5. Touch
Done
to return to the Home screen.
Optional fields
There are no optional fields for Fill Module Configuration.
Figure 3.38. Fill Control screen.