![ThermoFisher Scientific thermo scientific imPULSE S.U.M. User Manual Download Page 21](http://html1.mh-extra.com/html/thermofisher-scientific/thermo-scientific-impulse-s-u-m/thermo-scientific-impulse-s-u-m_user-manual_1105607021.webp)
Thermo Scientific
imPULSE S.U.M. with Touchscreen Console User’s Guide
|
16
Chapter 2
|
Hardware installation
2.1 Site preparation
2.1.1 Electrical connections
The power cable for the Touchscreen Console can be ordered
with a country-specific plug to connect to facility power. For some
countries, including the United States, the plug is amperage-specific.
Verify that your facility power matches the power requirements of the
Touchscreen Console power cable before use. The imPULSE S.U.M.
hardware cannot be used on circuits equipped with ground fault circuit
interrupter (GFCI) circuit protection, due to the potential for nuisance
tripping.
2.1.2 Hardware preparation
The hardware is shipped directly from the manufacturer and arrives
with various safety mechanisms in place. Follow the provided
guidelines to set up the imPULSE S.U.M. upon arrival.
CAUTION:
Any procedures that require the Touchscreen Console
to be open must be performed with the main electrical disconnect
in the locked out position, and all power sources removed from the
Touchscreen Console. For operator safety, secure the location of the
S.U.M. hardware by disabling the swivel casters before servicing.
2.1.3 Recommended tools and resources
For large vessels (500 L or larger), nylon slings are recommended to
use as lifting straps;
do not use chains or cables.
For fixed mounted
units, we recommend implementing a lifting and moving plan by a
professional rigging organization prior to lifting or moving the unit.
The imPULSE S.U.M. hardware will arrive with the following:
• Outer support container, including platform, tank, probe door
plates, Touchscreen Console (attached), electrical enclosure, and
an optional PCD (standard on 2,000 L or larger imPULSE mixers)
• Communication cables, with tool for connecting cables to the
Touchscreen Console (Figure 2.1)
• Equipment turnover package (ETP), located on a USB drive and
shipped separately
• (Optional) BPC hoist system for vessels 500 L and larger.