Thermo Scientific
imPULSE S.U.M. with Touchscreen Console User’s Guide
|
137
Chapter 6
|
Operating information
4. When approximately 3–5 liters remain in the BPC, lift the BPC from
the top. Hold the bottom drain line near the floor while lifting to
facilitate draining the final liter of liquid.
6.8 Shutdown and disposal
6.8.1 Shutting down the imPULSE S.U.M.
1. After draining the imPULSE S.U.M., verify that the motor agitation
is off, and turn off the power to the outer support container by
switching off the main power disconnect.
2. If the imPULSE S.U.M. hardware has been in contact with
caustic materials, rinse the affected areas with a light water rinse,
followed by a routine cleaning. See Chapter 7—Maintenance and
Troubleshooting for more information.
3. Return any loose items, such as tools, to their storage locations to
prevent accidental damage.
4. Remove the BPC from the outer support container and dispose of
the BPC according to your facility regulations (See section 6.8.3).
6.8.2 Decommissioning the Touchscreen Console
1. Turn off the AC input switch and the internal AC switch.
2. Disconnect the input power cord.
3. Turn off the AC output switch.
4. Disconnect the M12 AC power cable and the motor cable. Then
disconnect all other M12 cables.
5. Disconnect the optional remote E-Stop, if present.
6. Use caution while loosening the mounting screw and dismount the
unit.
7. Keep the unit in a proper storage space after it has been removed.