10 | Operation
SpeedVac™ SPD130DLX
HyperTerminal
Configuration
1. Power up the host computer and close any running
applications.
2. Open the HyperTerminal application by clicking on
Start
\
“Programs” \ “Accessories” \”Communications” \
“HyperTerminal.”
3. In the “Connection Description” box, enter the name
“SPD130DLX” and choose an icon and click
OK
.
4. Connect to the virtual COM port that is linked with the
SpeedVac™ concentrator attached to the computer
(USB virtual com port).
5. In the “COM Port Properties” box \ “Port Settings” folder
select the following options:
Bits per second: _ 115200
Data bits: _ 8
Parity: _ None
Stop bits: _ 1
Flow control _ None
After verifying the above settings, click
OK
.
6. In the main dialog box click on
File
\
Save
.
7. Exit the program by clicking on
File
\
Exit
\
Yes
.
8. Verify the program was saved by going to
Start
\
“Programs” \ “Accessories” \ “Communications” \
“HyperTerminal” \ “SPD130DLX”.
9. This completes the configuration of HyperTerminal.
10. Turn on SpeedVac™ system and connect to the virtual
COM port that is linked with the SpeedVac™ attached to
the computer.
11. Start HyperTerminal by clicking on
SPD130DLX
.
12. Initiate the run and SpeedVac™ will screen print
<temperature>, <heat time>, <run time>, <vacuum
pressure>
Lid Stay Mechanism
The unit is equipped with a lid-stay mechanism that will assist
the user in opening and closing the cover of the unit. It will
secure the lid and maintain it open, allowing the user to
operate the unit with both hands.
Figure 3. Lid stay mechanism view
Other Topics
Chamber Cover
. The cover must be closed before
beginning a run. If you press
Manual Run
or
Auto Run
and
the cover is open, the display shows “lid” to remind you to
close the cover. The run will start immediately after closing the
cover.
If a user tries to open the lid at the start of the run, when the
lid has already been locked, the unit will automatically abort
the run and signal the user with the “Lid” message on the
display and an error tone. Otherwise, during a run the user
cannot open the cover due to vacuum in the concentration
chamber. The cover is locked down at all times during a run
and whenever power to the unit is interrupted.
The cover lock is an additional safety feature that reduces the
risk of injury or damage from the spinning rotor.
Do NOT
bypass the cover lock to conduct a run with the cover open.
If removal of samples from the concentrator chamber during
a power failure is required, remove the screw holding the
cover lock release cord and lightly pull it while lifting the cover.
The cover can then be opened.
Figure 4. Opening chamber cover in emergency
WARNING:
Device lid can crush your fingers.
Keep away:
Do not reach between the device
and lid when opening or closing the device lid.