ThermoFisher Scientific SpeedVac SPD130DLX Installation And Operation Manual Download Page 11

8  |   Operation

SpeedVac™ SPD130DLX

display will revert temporarily to set points for approximately 5 
seconds.

Auto Run - Time 

1. Refer to the 

Manual Run

 section for start-up. 

2. To execute an Automatic time-based run, execute the 

following steps: 

a. Select the Run Type as “Time”.

b. Use the 

Select

 button and the “

+

”/ “

-

” buttons to 

select and modify the Temperature, “Heat Time” and 
“Run Time” parameters. The Run and Heat Time can 
be set from 0.01 to 9.59 hours (Heat Time also has 
“CCC” for continuous use). 

c. Select a vacuum set point: Since this is a time-based 

run, the vacuum will be the maximum allowed by the 
attached external vacuum pump, regardless of the 
vacuum pressure set point.

d. Press the 

Auto Run

 button to start the run. The 

cover locks and the lid locked indicator is lit. The rotor 
starts spinning. The run time display is counting 
down in 1 minute Intervals. The heat time is counting 
down if the set point is not set to “CCC” and if the 
temperature set point is not set to “no” (use the 
select button to view “Heat Time”). The temperature 
rises in 1 °C increments to set temperature. The 
vacuum is applied to the chamber and the vacuum 
pressure begins decrementing down from “HPr” 
(atmospheric pressure).

Note: 

If the cover is not closed, the display will show 

“Lid” and the run will not start. 

e. Press 

RL On/Off

 for radiant heat. Press at any time 

to turn ON and OFF. (As long as there is time left in 
the heat timer and if the temperature set point is not 
set to “no”).

f.

Once the time expires, the run will automatically stop, 
the display will show “End”, the SAV valves will click, 
also allowing air to bleed into the chamber. The unit 
will sound 1 beep that will be repeated until the user 
acts on it. 

g. After the rotor stops spinning, the cover unlocks and 

the lid locked indicator is turned off. The display 
reverts to last set parameters. 

h. Open the cover and remove samples.

Auto Run - Vacuum

1. Refer to the “Manual Run” section for start-up. 

2. To execute an Automatic Vacuum run, execute the 

following steps:

a. Select the Run Type as “Vac”

b. Use the 

Select

 button and the “

+

”/ “

-

” buttons to 

select and modify the Temperature, “Heat Time and 
“Run Time” parameters. The Run and Heat Time can 
be set from 0.01 to 9.59 hours (Heat Time also has 
“CCC” for continuous use). 

c. Select a vacuum set point at which the run should be 

terminated, ensure the attached external vacuum 
pump can generate the required vacuum level

d. Press the 

Auto Run

 button to start the run. The 

cover locks and the lid locked indicator is lit. The rotor 
starts spinning. The run time display will count up in 1 
minute intervals. The heat time is counting down if 
the set point is not set to “CCC” and if the 
temperature set point is not set to “no” (use the 

Select

 button to view “Heat Time”). The temperature 

rises in 1 °C increments to set temperature. The 
vacuum is applied to the chamber and the vacuum 
pressure begins decrementing down from “HPr” 
(atmospheric pressure).

Note: 

If the cover is not closed, the display will show 

“Lid” and the run will not start. 

e. Press 

RL On/Off

 for applying radiant heat. Press at 

any time to turn ON and OFF. (As long as the heat 
timer has not expired, and if the temperature set 
point is not set to “no”).

f.

Once the unit senses the set vacuum pressure in the 
chamber has been achieved, the run will 
automatically stop and the display will show “End”. 
The valve will click, allowing air to bleed into the 
chamber. The unit will sound 1 beep that will be 
repeated until the user acts on it. 

g. After the rotor stops spinning, the cover unlocks and 

the lid locked indicator is turned off. The display 
reverts to last set parameters. 

h. Open the cover and remove samples.

Drying Rate

Thermal energy can be applied to the concentrator chamber 
to counteract the cooling effect of evaporation, maintain the 
samples in the liquid state, and accelerate the concentration 
run. Select the highest drying rate if you desire accelerated 

Summary of Contents for SpeedVac SPD130DLX

Page 1: ...Installation and Operation 80302133 Revision A July 2018 SpeedVac SPD130DLX Vacuum Concentrator...

Page 2: ...mo Fisher Scientific Inc provides this document to its customers with a product purchase to use in the product operation This document is copyright protected and any reproduction of the whole or any p...

Page 3: ...4 Installation of SPD130DLX 4 Operation 5 Description Of Control Panel 5 Manual Run 7 Auto Run Time 8 Auto Run Vacuum 8 Drying Rate 8 Preset Program Settings 9 Exporting Live Run Data 9 HyperTerminal...

Page 4: ...th dual timer for automatic or manual control of run conditions it also features preset and user defined programs which enable user to retrieve and store application protocols for future use The compa...

Page 5: ...dicates extreme low temperatures and high risk of frostbite Do not touch bare metal or samples with unprotected body parts This symbol indicates possible pinch points which may cause personal injury T...

Page 6: ...tive Power 115 VAC 60 Hz 6 amps 230 VAC 50 Hz 3 amps Vacuum Chamber TEFLON coated chemical resistant aluminum casting Cover Heated cover with radiant lamps includes a safety interlock Induction Motor...

Page 7: ...130DLX units configured for 230 VAC 50 Hz should be plugged into a circuit rated for at least 3 amps Note Do not use any detachable power cord that is not adequately rated for the unit Installation of...

Page 8: ...hand to ensure alignment of the pin with the groove Secure the assembly by screwing the retaining knob into the drive shaft above the rotor Tighten it firmly but not excessively Description Of Contro...

Page 9: ...meters is shown by the corresponding indicator that is illuminated 8 Minus Decrements the value of the selected parameter 9 Plus Increments the value of the selected parameter 10 View During a run pre...

Page 10: ...tor with the supplied knob hand tight Close cover 10 Pre heat may be selected at this time to apply a 45 C heat to the chamber 11 Press the Manual Run button The cover locks and the lid locker indicat...

Page 11: ...and remove samples Auto Run Vacuum 1 Refer to the Manual Run section for start up 2 To execute an Automatic Vacuum run execute the following steps a Select the Run Type as Vac b Use the Select button...

Page 12: ...C The heat time is expressed as a 3 digit integer in minutes or CCC for continuous heating The runtime is expressed as a 3 digit integer in minutes The vacuum pressure is represented as a floating poi...

Page 13: ...id stay mechanism that will assist the user in opening and closing the cover of the unit It will secure the lid and maintain it open allowing the user to operate the unit with both hands Figure 3 Lid...

Page 14: ...cid Ammonium hydroxide n butyl alcohol Chloroform Cyclohexane DCM Ethyl acetate Ethanol Ethyl ether n Hexane n Heptane Isopropyl alcohol Isobutyl alcohol Methanol Petroleum ether n propyl alcohol Pyri...

Page 15: ...0mm RH32 13 15 8 17 5 X 102 mm 15ml Corex tubes RH8 17 5 10 8 18 X 100 mm 17 x 95 16 x 100 RH8 18 CENTRIFUGE TUBES 12 10 15ml conical 16 x 120 mm RH10 15 40 6 50ml conical 28 x 115 mm RH6 50 FLASKS 80...

Page 16: ...neutralizing solution for Oligo Preps Used on UVS850DDA only SCT120 Chemical trap order cartridges separately DTK120R Chemical trap kit for radioactivity compatible with all Thermo vacuum concentrator...

Page 17: ...Pump ANT100 Exhaust recovery Post trap Vacuum Tubing From Cold Trap to Back of SPD130DLX UVS850DDA Vacuum Tubing From Back of SPD130DLX To Cold Trap SpeedVac Concentrator SPD130DLX Vacuum Tubing RVT45...

Page 18: ...MA number issued by Thermo Costs of shipping the unit are not covered under warranty The warranty obliges you to follow the precautions in this manual When returning apparatus that may contain hazardo...

Page 19: ...riciclato tramite queste medesime Ulteriori informazioni sulla conformit di Thermo Fisher Scientific con queste Direttive l elenco delle ditte di riciclaggio nel Vostro paese e informazioni sui prodo...

Page 20: ...30 France 33 2 2803 2180 33 2 2803 2180 Germany 0800 1 536 376 49 6184 90 6000 0800 1 536 376 Italy 32 02 95059 552 39 02 95059 552 432 254 375 Netherlands 31 76 579 55 55 31 76 571 4440 Nordic Baltic...

Reviews: