1.
In the home screen, tap
(Settings)
4
Maintenance and
service
4
Calibration
4
Status and reminders
.
The
Status and Reminders
screen is displayed. It shows the following
information:
• Type of calibration
• Status of the calibration
• Last calibration
• Date that the calibration expires
• If a reminder is set
2.
Tap a calibration row, then tap
View
to view the history of that specific
calibration type.
• Date performed
• Result of the calibration
• Whether the calibration is current
• Operator
3.
Tap
Cancel
to return to the
Status and Reminders
screen.
Set calibration frequency and reminders on the instrument
1.
In the home screen, tap
(Settings)
4
Maintenance and
service
4
Calibration
4
Status and reminders
.
The
Status and Reminders
screen is displayed.
2.
Tap a calibration row, then tap
Edit
.
3.
Tap on the
Expiration Interval
field, then enter the number of days.
The default is 730 days (2 years).
4.
Tap on the
Reminder
field, then enter the number of days
The
Next Reminder
field will be populated.
5.
Set the
On/Off
slider to the
On
position to enable a reminder about calibration.
6.
Tap
Save
.
The calibration reminder can be set as a home screen notification (see “Enable and
disable home screen notifications” on page 168).
View calibration status and set reminders on the Connect platform
Note:
The calibration reminders feature requires a connection between the
instrument and a computer network.
1.
Sign in to Connect on a browser.
2.
Tap
(Instrument)
to open InstrumentConnect.
3.
Select any of your registered instruments.
Chapter 7
Calibrate and verify instrument performance
View the calibration status and set reminders
7
82
QuantStudio
™
6 Pro and 7 Pro Real-Time PCR Systems User Guide