Delete plate files
System templates cannot be deleted. A plate file that was saved from a system
template can be deleted.
For descriptions of files, see “Files” on page 27.
1.
In the home screen, tap
Load plate file
or
Set up run
.
2.
Tap the appropriate folder.
•
Run Queue
•
Public
•
Connect
•
Network Drive
•
USB
3.
Tap
Manage
.
4.
Tap the file, then tap
Delete files
.
Multiple files can be selected.
5.
Tap
Yes
to confirm.
Tap
(Back)
to return to the home screen.
Manage data files
View QC Check status after a run
The status dial displays
Run complete
when a run is complete. After
Done
is
tapped, the amplification plot can no longer be viewed on the instrument.
1.
In the home screen after a run is complete, see the status of the QC Check.
The status of the QC Check is displayed under the status dial.
Chapter 6
View and manage files
Manage data files
6
QuantStudio
™
6 Pro and 7 Pro Real-Time PCR Systems User Guide