Pause an instrument run
Pause a run to add reagents to the plate or to edit the number of cycles.
1.
In the home screen during a run, tap
(Actions)
.
2.
Tap
Pause run
, then enter a pause temperature.
3.
(Optional)
Tap
Edit
to change the number of cycles.
4.
(Optional)
Tap
(Eject)
to open the drawer, then tap
(Eject)
to close the
drawer.
CAUTION!
PHYSICAL INJURY HAZARD.
During instrument operation, the
plate temperature can reach 100°C. If you want to access the plate during a run
pause, enter room temperature as the pause temperature and allow the plate to
cool to room temperature before handling.
Stop an instrument run
CAUTION!
PHYSICAL INJURY HAZARD.
During instrument operation, the plate
temperature can reach 100°C. If you want to access the plate during a run pause,
enter room temperature as the pause temperature and allow the plate to cool to room
temperature before handling.
1.
In the home screen during a run, tap
(Actions)
.
2.
Tap
Stop run
.
3.
Unload the plate (see “Unload a plate from the instrument” on page 51).
Chapter 5
Options for a run
Pause an instrument run
5
QuantStudio
™
6 Pro and 7 Pro Real-Time PCR Systems User Guide
67