Figure 25 Start run pane
1
Tip box status
2
Section of sample plate to extract samples
3
Confirm foil is removed from the selected plate
checkbox
4
OpenArray
™
Plate position (color corresponds to the section of the sample plate)
5
TPF or SPF file name or serial number of the OpenArray
™
Plate
6
Auto-fill tip boxes
button
Before you begin—full run workflow
• Prepare samples (DNA and PCR reaction mix) in a 384-well plate, according to the user guide or
the application guide for your assay.
• Distribute the samples according to the layout determined in the
Map Plates
workflow.
• Place the sample plate in the sample plate holder on the instrument deck, with the notch to the left.
Do not stack sample plates.
• Load the tip boxes, then remove the tip box covers.
Do not stack the tip boxes.
• Place the OpenArray
™
Plates in the plate holders.
• Clear the instrument deck, empty and replace the waste bin, and close the instrument door.
For more information on setting up the instrument deck, see “Set up the system”
.
Chapter 7
Set up and start a run with the full run workflow
Before you begin—full run workflow
7
50
QuantStudio
™
12K Flex OpenArray
™
AccuFill
™
System User Guide