background image

 

22 

Chapter 3 | Setup Menu 

3.2.15 Set Date Value 

Set the month, day and year values. The date format used 
for this Setup Menu item will match the setting in the Set 
Date Format item. For this example, the month-day-year 
(MM.DD.YYYY) format is shown. 

 

Press the 

 or 

 key to scroll through the Setup 

Menu list until the Set Date Value item is shown. 

 

 

The month value will flash. Press the 

 or 

 key to 

adjust the month.  

a.

 

Press the 

 key to increase by one month. 

Press and hold the 

 key to quickly increase. 

b.

 

Press the 

 key to decrease by one month. 

Press and hold the 

 key to quickly decrease. 

 

Press the “Enter” key to save the change. 

 

 

The day value will flash. Press the 

 or 

 key to 

adjust the day.  

a.

 

Press the 

 key to increase by one day. Press 

and hold the 

 key to quickly increase. 

b.

 

Press the 

 key to decrease by one day. 

Press and hold the 

 key to quickly decrease. 

 

Press the “Enter” key to save the change. 

 

 

The year value will flash. Press the 

 or 

 key to 

adjust the year.  

a.

 

Press the 

 key to increase by one year. Press 

and hold the 

 key to quickly increase. 

b.

 

Press the 

 key to decrease by one year. 

Press and hold the 

 key to quickly decrease. 

 

Press the “Enter” key to save the change. 

 

3.2.16 Set Time Value 

Set the time in AM/PM or 24 hour formats. 

 

Press the 

 or 

 key to scroll through the Setup 

Menu list until the Set Time Value item is shown. 

 

 

The AM/PM setting will flash. Press the 

 or 

 key 

to set the AM or PM time or to select the 24 hour 
format. 

 

Press the “Enter” key to save the change. 

 

The hour value will flash. Press the 

 or 

 key to 

adjust the hours.  

a.

 

Press the 

 key to increase by one hour. 

Press and hold the 

 key to quickly increase. 

b.

 

Press the 

 key to decrease by one hour. 

Press and hold the 

 key to quickly decrease. 

 

Press the “Enter” key to save the change. 

 

The minutes value will flash. Press the 

 or 

 key 

to adjust the minutes.  

a.

 

Press the 

 key to increase by one minute. 

Press and hold the 

 key to quickly increase. 

b.

 

Press the 

 key to decrease by one minute. 

Press and hold the 

 key to quickly decrease. 

 

Press the “Enter” key to save the change. 

 

Summary of Contents for Orion Lab Star EC112

Page 1: ...Operations Manual Electrochemistry Meters 68X002223 Revision 01 June 2022 Orion Lab Star EC112 Conductivity Meter User Manual...

Page 2: ......

Page 3: ...u Value Changes 9 2 3 3 Mode Shortcut Key 9 2 3 4 Date and Time View Shortcut Key 9 2 4 Display Overview 10 2 5 Meter Connections 11 2 6 Electrode Information 11 2 7 Meter Startup Sequence 11 Chapter...

Page 4: ...24 Chapter 4 25 Calibration 25 4 1 Conductivity Calibration 25 4 1 1 Calibration Overview 25 4 1 2 Conductivity Calibration Procedure 26 4 1 3 Conductivity Calibration Example 26 4 1 4 Conductivity Va...

Page 5: ...fications 35 7 2 Meter Dimensions 36 7 3 Warranty 36 7 3 1 Return of Items 36 Chapter 8 37 Regulatory Compliance 37 8 1 European Union 37 8 2 Product Safety 37 8 3 Electromagnetic Compatibility 37 8 3...

Page 6: ...d result in minor to moderate injury or damage to the equipment Note This symbol in the context of a Note indicates to follow the manufacturer specified instructions notes and requirements set out in...

Page 7: ...on a clean dry surface Identify the side of the meter that the holder will be installed on and remove the screw between the circles on that side of the meter Align the base plate of the holder with th...

Page 8: ...ack the power supply provided with the meter Select the appropriate wall plug plate for the power outlet that will be used Slide the appropriate wall plug plate into the groove on the back of the powe...

Page 9: ...up Menu Press to accept a calibration point in the Calibration mode Press to power on the meter When the meter is on press to turn the backlight off or on When the meter is on press and hold for three...

Page 10: ...y Icon Indicates meter is in the Setup Menu and viewing the data log or calibration log Log Saved Icon Indicates meter is actively saving data to the data log or calibration log Continuous Icon Indica...

Page 11: ...ns for specific care and maintenance procedures The following are general recommendations for electrode preparation Remove the packaging from the electrode and save for storage Rinse the electrode usi...

Page 12: ...mperature Input When conductivity electrode with built in ATC sensor is connected use the Temp Calibration menu to perform a temperature offset calibration When conductivity electrode without ATC sens...

Page 13: ...ta and calibration log options Press key Press key Press key Press key Press key Press the Enter key to view the display log 3 2 1a Viewing the Data Log Press the Enter key to view the display log For...

Page 14: ...to view the display log For the calibration log press the or key to change the view from the date time screen to the calibration data screen Press or key For the calibration log press the or key to sc...

Page 15: ...off Press the Enter key to save the change 3 2 3 Export All Logs Send all saved data logs and calibration logs to the selected export device Note that this menu will not be shown if the Log Export Typ...

Page 16: ...he temperature offset value will be shown in the 2nd field a Press the key to increase the value by 0 1 Press and hold the key to quickly increase b Press the key to decrease the value by 0 1 Press an...

Page 17: ...ce reading Once a calibration is performed the entered nominal cell constant K multiplied by the calculated calibration factor is used to determine conductivity salinity and TDS measurement values fro...

Page 18: ...tup Menu list until the Temperature Compensation Coefficient item is shown Press the or key to set the coefficient value a Press the key to increase the value by 0 01 Press and hold the key to quickly...

Page 19: ...ly updated on the display and the stopwatch stabilizing or checkmark stable icon indicates the measurement stability status Press the Log Export key to save a measurement to the data log and export Au...

Page 20: ...me value by one second Press and hold the key to quickly decrease Press the Enter key to save the change 3 2 12 Calibration Due Alarm Set the calibration due alarm interval from 0 hours off to 168 hou...

Page 21: ...mperature Units item is shown Press the or key to set the temperature units Press the Enter key to save the change 3 2 14 Set Date Format Set the date format as month day year MM DD YYYY or day month...

Page 22: ...lash Press the or key to adjust the year a Press the key to increase by one year Press and hold the key to quickly increase b Press the key to decrease by one year Press and hold the key to quickly de...

Page 23: ...nter key to save the change 3 2 18 Sleep Mode Set the sleep mode on or off When the meter sleep mode is on the meter will enter sleep mode when no keys are pressed for 20 minutes Once the meter is in...

Page 24: ...erased The meter will confirm when the selected log is fully erased The meter will return to the measurement screen 3 2 20 Factory Reset Erase all data logs calibration logs and settings and return th...

Page 25: ...25 C The Temperature Compensation Coefficient setting in the Setup Menu will determine how the conductivity calibration standards are displayed during the calibration If temperature compensation is se...

Page 26: ...Proceed to the next calibration point or save and end the calibration a To proceed to the next calibration point repeat steps 3 5 or b To save and end the calibration press the Cal key The meter will...

Page 27: ...value is shown and then the Calibration Done screen is shown The meter will proceed to the main measure mode 4 1 4 Conductivity Value vs Temperature The following table lists theoretical values for t...

Page 28: ...4 3 TDS Calibration Ensure the active measure mode is TDS Press the Cal key to start the calibration Rinse the conductivity electrode and place into the TDS standard Wait for the TDS value to stabili...

Page 29: ...temperature and NIST traceable thermometer Press the Menu key to enter the Setup Menu Press the or key to scroll through the Setup Menu list until the Temp Calibration item is shown The temperature v...

Page 30: ...e entire time the meter is in the measurement mode 5 2 Conductivity Salinity and TDS Measurements Rinse the electrode with distilled water or appropriate solution and blot gently with a lint free tiss...

Page 31: ...erature meter Both the Actual TDS and the Actual Conductivity values must be in the same magnitude of units For example if the TDS value is in ppm the conductivity value must be in S cm if the TDS val...

Page 32: ...er using computer programs such as LIMs Putty LabView HyperTerminal or similar programs a To record the COM port location of the computer cable use the computer s Device Manager tool 5 7 2 Remote Comm...

Page 33: ...rograms and Features tool Make sure the meter is interfaced with the computer using the USB computer cable and the driver has been installed according to the instructions in the USB Computer Cable Int...

Page 34: ...the sample solution with the electrode sensing plates and temperature sensor submerged below the top of the sample solution o Verify cables do not have any damage o Perform a factor reset on the mete...

Page 35: ...oint Offset adjustment Up to 5 C Features Display 5 backlit LCD Time and date Yes Time format 12 hour AM PM or 24 hour Read types Continuous Auto Read Timed Timed interval 5 seconds to 60 minutes Data...

Page 36: ...d on a charged basis The warranty on your meter shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the en...

Page 37: ...environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio commun...

Page 38: ...der entsorgt werden Mehr Informationen ber die Einhaltung dieser Anweisungen durch Thermo Scientific dieVerwerter und Hinweise die Ihnen n tzlich sein k nnen die Thermo Fisher Scientific Produkte zu i...

Page 39: ...TARA 109 9 2 Electrode and Solution Ordering Info Description Catalog Number Orion conductivity ATC electrode 2 cell K 1 0 epoxy platinum 1 5m cable 011050MD Orion conductivity ATC electrode 2 cell K...

Page 40: ...ety today Whether online in the plant in the field or in the lab scientists trust Orion products deliver precise and accurate readings which help you keep our communities healthier cleaner and safer F...

Reviews: