background image

 

 

Nucleic Acid Applications

Measure dsDNA, ssDNA or RNA

Thermo Scientific

NanoDrop Eight User Guide         19

Nucleic Acid Reported Results

dsDNA measurement screen

For each measured sample, the dsDNA, ssDNA and RNA applications show the UV 
absorbance spectrum and a summary of the results. Below is an example of the 
measurement screen of the instrument software:

Click row to select sample 
and update spectrum.

UV spectrum

Right click graph 
area to view

 display 

options

.

Measure sample

End experiment

Run Blank

Purity ratios

Menu of table 
options;
click to choose 
which columns to 
report

Nucleic acid 
concentration

Sample name; 
select to edit

Application

Well plate 
options

Baseline 
correction

Wellplate view; 
toggle to view/ hide 
well plate map

Load Sample ID file

View history data

Auto measure

Note  

Micro-volume absorbance measurements are normalized to a 10.0 mm 

pathlength equivalent.

Summary of Contents for NanoDrop Eight

Page 1: ...NanoDrop Micro UV Vis Spectrophotometer NanoDrop Eight User Guide M020 NanoDrop Eight UG Revision Date December 2021...

Page 2: ...l technical information in this document is for reference purposes only System configurations and specifications in this document supersede all previous information received by the purchaser Thermo Fi...

Page 3: ...se contact 12 Chapter 3 Application Measurement Ranges 13 Detection Limits for All Applications 13 Chapter 4 Nucleic Acid Applications 15 Measure dsDNA ssDNA or RNA 15 Measure dsDNA ssDNA or RNA 15 Be...

Page 4: ...ein A280 Measurements 51 Calculations for Protein A280 Measurements 51 Chapter 6 Custom Applications 57 Measure UV Vis 58 Measure UV Vis 58 Best practices for UV Vis measurements 59 UV Vis Reported Re...

Page 5: ...rement Screen Display Options 124 Chapter 8 Maintenance 127 Maintenance Schedule 128 Daily Maintenance 128 Periodic Maintenance 128 Every 6 Months 128 Maintaining the Pedestals 129 Cleaning the Pedest...

Page 6: ...Contents vi NanoDrop Eight User Guide Thermo Scientific...

Page 7: ...thout the need for dilutions The instrument accurately measures up to 8 individual 1 l samples in one measurement cycle The software allows the user to measure samples using either the full 8 channel...

Page 8: ...r features the patented micro volume sample retention system USB B port Connect the NanoDrop Eight to your PC via USB to operate the instrument using the PC software Note Before operating a NanoDrop E...

Page 9: ...aining plastic consumables and liquid photometric standard used to check instrument performance For more information see Performance Verification Specially formulated conditioning compound that can be...

Page 10: ...e Spectrophotometer Instrument Detection Limits 10 NanoDrop Eight User Guide Thermo Scientific Instrument Detection Limits Pathlength mm Upper Detection Limit 10 mm Equivalent Absorbance 1 0 12 5 0 2...

Page 11: ...com nanodropsw On the website locate the Register Your Instrument button and follow the instructions to register the instrument Computer Requirements Required Windows Operating System Windows 10 Enter...

Page 12: ...llation folder downloaded from the internet 2 Launch Start exe and click the Software button The software installer will run Technical Support For U S Canada Support please contact Thermo Fisher Scien...

Page 13: ...f 0 04 200 A Sample Type Lower Detection Limit Upper Detection Limit Typical Reproducibilitya a Based on five replicates SD ng L CV dsDNA 2 0 ng L 10 000 ng L 2 0 ng L for sample concentrations betwee...

Page 14: ...3 Application Measurement Ranges 14 NanoDrop Eight User Guide Thermo Scientific This page is intentionally blank...

Page 15: ...A230 A single point baseline correction can also be used To measure dsDNA ssDNA or RNA samples Measures the concentration of purified dsDNA ssDNA or RNA samples that absorb at 260 nm Measure dsDNA ssD...

Page 16: ...ify a baseline correction if desired 3 Pipette 1 2 L blanking solution onto the lower pedestal and lower the arm 4 Select Blank and wait for the measurement to complete Tip If Auto Blank is On the bla...

Page 17: ...s Isolate and purify nucleic acid samples before measurement to remove impurities Depending on the sample impurities could include DNA RNA free nucleotides proteins some buffer components and dyes See...

Page 18: ...ring a Blank for more information For micro volume measurements Ensure pedestal surfaces are properly cleaned and conditioned If possible heat highly concentrated or large molecule samples such as gen...

Page 19: ...trument software Click row to select sample and update spectrum UV spectrum Right click graph area to view display options Measure sample End experiment Run Blank Purity ratios Menu of table options c...

Page 20: ...minant corrected Settings for Nucleic Acid Measurements Baseline Correction For the dsDNA ssDNA or RNA enter baseline correction wavelength in nm or use default value 340 nm Optional user defined base...

Page 21: ...factor f 1 b where wavelength dependent molar extinction coefficient in ng cm L b sample pathlength in cm As a result analyte concentration c is calculated as c A 1 b or c A f where c analyte concent...

Page 22: ...on Measured Values Note For micro volume absorbance measurements the spectra are normalized to a 10 mm pathlength equivalent A260 absorbance Nucleic acid absorbance values are measured at 260 nm using...

Page 23: ...orbance at 230 nm An A260 A230 purity ratio between 1 8 and 2 2 is generally accepted as pure for DNA and RNA Note Although purity ratios are important indicators of sample quality the best quality in...

Page 24: ......

Page 25: ...tal measurements with the NanoDrop Eight instrument lift the instrument arm and clean the upper and lower pedestals At a minimum wipe the pedestals with a new laboratory wipe For more information see...

Page 26: ...k measurement starts automatically after you lower the arm 5 Lift the arm and clean all pedestals with a new laboratory wipe 6 Pipette 1 2 L sample solution onto the pedestals and lower the arm 7 Star...

Page 27: ...oom Right click and select Autoscale to fit spectra to window Measure sample solution Select to toggle Auto Measure ON OFF Sample names select to add or edit Run blank Back to Home screen Tips Click s...

Page 28: ...aw equation Based on extinction coefficient and pathlength f 1 260 b where f factor molar extinction coefficient at 260 nm in ng cm L b sample pathlength in cm 1 cm for nucleic acids measured with the...

Page 29: ...imits are dependent on the upper absorbance limit of the instrument and the user defined extinction coefficients To calculate upper detection limits for nucleic acid samples To calculate upper detecti...

Page 30: ...4 Nucleic Acid Applications 30 NanoDrop Eight User Guide Thermo Scientific This page is intentionally blank...

Page 31: ...anoDrop Eight instrument lift the instrument arm and clean the upper and lower pedestals At a minimum wipe the pedestals with a new laboratory wipe For more information see Cleaning the Pedestals To m...

Page 32: ...ift the arm and clean all pedestals with a new laboratory wipe 6 Pipette 1 2 L sample solution onto the pedestal and lower the arm 7 Start the sample measurement If Auto Measure is On lower arm if Aut...

Page 33: ...ed results Here is an example Oligo DNA and Oligo RNA reported values Nucleic acid concentration Click row to select sample and update spectrum click more rows to overlay spectra UV spectrum Purity ra...

Page 34: ...ate sample measurement was taken nucleic acid concentration A260 A280 A260 A230 A260 A280 factor oligo sequence baseline correction Related Topics Basic Instrument Operations Oligo Calculations Note T...

Page 35: ...ence for RNA Use the G A U and C keys to specify the RNA base sequence Specify your DNA or RNA base sequence click the corresponding keys You can also enter base sequence using the keyboard or by copy...

Page 36: ...ng L Molecular Weight of oligo calculated from user defined base sequence Number of Bases entered Molar Ext Coefficient 260 nm Molar extinction coefficient of oligo in ng cm L at 260 nm calculated fro...

Page 37: ...sentially randomized sequences but not for short synthetic oligo sequences To ensure the most accurate results we use the exact value of 260 to calculate oligonucleotide concentration The NanoDrop sof...

Page 38: ...A260 absorbance Note For micro volume absorbance measurements and measurements taken with nonstandard other than 10 mm cuvettes the spectra are normalized to a 10 mm pathlength equivalent Nucleic aci...

Page 39: ...cally have a higher 260 280 ratio due to the higher ratio of Uracil compared to that of Thymine Reported Values Nucleic acid concentration Reported in selected unit i e ng L g l or g mL Calculations a...

Page 40: ...4 Nucleic Acid Applications Measure Oligo DNA or Oligo RNA 40 NanoDrop Eight User Guide Thermo Scientific...

Page 41: ...and one absorbance ratio A260 A280 A single point baseline correction can also be used This application does not require a standard curve To measure Protein A280 samples Measures the concentration of...

Page 42: ...desired 3 Pipette 1 2 L blanking solution onto the lower pedestal and lower the arm 4 Click Blank and wait for the measurement to complete Tip If Auto Blank is On the blank measurement starts automat...

Page 43: ...on the sample impurities could include DNA RNA and some buffer components See Preparing Samples for more information Ensure the sample absorbance is within the instrument s absorbance detection limit...

Page 44: ...r or application See Choosing and Measuring a Blank for more information For micro volume measurements Ensure pedestal surfaces are properly cleaned and conditioned Proteins tend to stick to pedestal...

Page 45: ...ine correction wavelength Click row to select sample and update spectrum UV spectrum Right click graph area to view display options Measure sample End experiment Run Blank Purity ratios Menu of table...

Page 46: ...select the settings icon to view Protein A280 Setup Protein A280 settings The Protein A280 application provides a variety of sample type options for purified protein analysis Each sample type applies...

Page 47: ...lculates BSA Bovine Serum Albumin protein concentration using mass extinction coefficient of 6 7 L gm cm at 280 nm for 1 i e 10 mg mL BSA solution Assuming MW is 66 400 daltons Da molar extinction coe...

Page 48: ...inction coefficient in L gm cm for 10 mg mL 1 protein solution a To add or edit a custom protein use Protein Editor Setting Available Options Mass Ext Coefficient L gm cm Description Setting Available...

Page 49: ...Protein Editor select to show the New Protein Type box 2 Enter a unique Name for the new protein 3 Enter a Description for the new protein 4 Specify whether to enter Molar Extinction coefficient or Ma...

Page 50: ...loDaltons kDa 5 Select Save to close the New Protein Type box The new custom protein appears in the Type list in Protein A280 Setup and Proteins Labels Setup Edit custom protein 1 In Protein Editor cl...

Page 51: ...10 For example if the sample s mass extinction coefficient at 280 nm is 6 7 for a 1 10 mg mL solution the equation looks like this 200 6 7 10 298 5 or 300 Calculations for Protein A280 Measurements T...

Page 52: ...shown at right for selecting an appropriate extinction coefficient for each measured sample to be used in conjunction with Beer s Law to calculate sample concentration If the extinction coefficient o...

Page 53: ...ctly The equation at the right shows the relationship between molar extinction coefficient molar and percent extinction coefficient 1 Conversions Between molar and 1 molar 10 1 MWprotein Example To de...

Page 54: ...sample such as residual solvents and reagents typically used during sample purification Measured Values A280 absorbance Note For micro volume absorbance measurements the spectra are normalized to a 1...

Page 55: ...e at 260 nm to corrected absorbance at 280 nm An A260 A280 purity ratio of 0 57 is generally accepted as pure for proteins Note Although purity ratios are important indicators of sample quality the be...

Page 56: ...5 Protein Applications Measure Protein A280 56 NanoDrop Eight User Guide Thermo Scientific...

Page 57: ...s Use the NanoDrop Eight to perform UV Vis measurements The UV Vis application allows the instrument to function as a conventional spectrophotometer Up to 40 wavelengths from 190 nm to 850 nm can be m...

Page 58: ...new laboratory wipe For more information see Cleaning the Pedestals To measure a sample using the UV Vis application 1 From the Home screen from the Custom tab select UV Vis 2 Specify up to 40 wavelen...

Page 59: ...w wipe Best practices for UV Vis measurements Ensure the sample absorbance is within the instrument s absorbance detection limits Blank with the same buffer solution used to re suspend the analyte of...

Page 60: ...o Home screen Tips Click sample row to select sample and update spectrum Shift click multiple sample rows to overlay spectra Click a sample and hover locations on spectra to view measurement values Sa...

Page 61: ...sample row to select sample and update spectrum Shift click multiple sample rows to overlay spectra Click a sample and hover locations on spectra to view measurement values Sample rows Select to add...

Page 62: ...nm This is the wavelength the software will use to determine the pathlength selection Automated Pathlength On or Off affects pedestal measurements only Optional automated pathlength selection Allows t...

Page 63: ...in Before taking measurements with the NanoDrop Eight instrument lift the instrument arm and clean the upper and lower pedestals At a minimum wipe the pedestals with a new laboratory wipe For more inf...

Page 64: ...lect Custom Methods 2 In the method selection pane select the method to run Information about the selected method appears in the method details pane 3 Select 4 Follow the on screen instructions to mea...

Page 65: ...thod reported values Result name Measurement Range Analysis wavelength correction Click row to select sample and update spectrum UV spectrum Right click graph area to view display options Measure samp...

Page 66: ...method to be used for sample measurements with user defined settings Create new custom method From the NanoDrop Home screen select the Custom tab and select Custom Methods In the Manage Custom Methods...

Page 67: ...and concentration of each standard and select the curve fit type Alternatively load a standard curve Enter or choose remaining custom settings as needed Choose Save Note Any errors in the method will...

Page 68: ...measurement screen opens with the standard curve loaded You can begin measuring samples View or edit custom method Select Custom Method existing methods are listed in Select Method box along with thei...

Page 69: ...orrection are used make sure your selected spectral range includes your specified baseline correction and or analysis correction wavelength For micro volume absorbance measurements and measurements ta...

Page 70: ...ficient for 1 cm pathlength and use adjacent drop down list to select appropriate unit Equation below shows how extinction coefficient is used to calculate sample concentration c A b where c analyte c...

Page 71: ...dictates that standards be measured from the lowest concentration of the standard analyte stock to the highest Concentration range of the standards must cover the dynamic range of the assay and the e...

Page 72: ...olume measurements only When Automated Pathlength is selected software selects the optimal pathlength between 1 0 mm and 0 1 mm based on sample absorbance at the analysis wavelength For example when s...

Page 73: ...orted in Data Table and Sample Details screens Unit Enter unit for reported result After a measurement the unit is reported in Data Table and Sample Details screens Edit Edit selected formula for curr...

Page 74: ...n custom method From the Custom Methods screen select a custom method Select Follow the on screen instructions to measure a sample Export Custom Method Export a custom method in order to run it and st...

Page 75: ...tings From Custom Methods screen select a custom method from the list of available methods From the drop down menu choose Edit Edit method settings as desired Choose Save Delete custom method From Cus...

Page 76: ...6 Custom Applications Measure Custom 76 NanoDrop Eight User Guide Thermo Scientific...

Page 77: ...ts Micro Volume Sampling How it Works 78 Set Up the Instrument 80 Measure a Micro Volume Sample 82 Sample Modes 86 Prepare Samples and Blanks 86 Basic Instrument Operations 92 Acclaro Sample Intellige...

Page 78: ...surface tension to hold a small volume of sample between two pedestals upper and lower The patented sample retention system enables the measurement of highly concentrated samples without the need for...

Page 79: ...nt is blanked a reference spectrum is taken of the blanking solution and stored in memory For each sample measurement the sample intensities along with the blank intensities are used to calculate the...

Page 80: ...e set up to apply a baseline correction to each measurement to minimize any offset caused by light scattering particulates in the sample spectra The correction subtracts the absorbance value at a refe...

Page 81: ...these specifications operating temperatures 5 C 35 C 41 F 95 F relative humidity non condensing 20 80 Locate the instrument away from air vents and exhaust fans to minimize evaporation After the inst...

Page 82: ...angle for optimal use of the pipette guide Use calibrated precision pipettor 0 2 L volume range with well fitting low retention precision tips to apply sample material to instrument for measurement I...

Page 83: ...Materials Recommended sample volumes To measure a micro volume sample Application Sample Volume Nucleic acid aqueous solution 1 La a Use 2 L for samples that contain materials that may reduce surface...

Page 84: ...ement to complete Tip If Auto Blank is On blank measurement starts automatically after you lower the arm Lift the arm and clean all pedestals with a new laboratory wipe 4 Measure the first sample Pipe...

Page 85: ...ring samples Select the End Experiment icon see previous image Enter an experiment name or leave the default experiment name Select Save Lift the arm and clean all pedestals with a new wipe If finishe...

Page 86: ...e area This on screen sample plate will populate with Sample ID information from an imported sample ID file You can also manually enter Sample ID or other identifying information The sample status col...

Page 87: ...before taking a measurement Avoid introducing bubbles when mixing and pipetting Choosing and Measuring a Blank The buffer used to resuspend a sample analyte can contribute absorbance Blanking minimiz...

Page 88: ...ow concentration samples See below for details Good blanking buffer measured abs 0 04 Problems associated with blanking Residual sample was left on pedestal before blank measurement was performed Resu...

Page 89: ...n rerun blanking cycle or measure new blank using new aliquot of appropriate buffer solution then measure new aliquot of unknown sample For most applications blank with the same buffer solution used t...

Page 90: ...u plan to use for sample analyses Supplies needed lint free laboratory wipes calibrated precision pipettor 0 2 L buffer solution for evaluation To run a blanking cycle For quick demonstration watch mu...

Page 91: ...lution onto the pedestal and lower the arm Start the sample measurement if Auto Measure is On lower arm if Auto Measure is off lower arm and select Measure Wait for measurement to complete The resulti...

Page 92: ...op Eight Home Screen NanoDrop Eight Measurement Screens View History NanoDrop Eight General Operations NanoDrop Eight Home Screen These operations are available from the NanoDrop Eight Home screen His...

Page 93: ...mation is described below Control options Serial number Instrument serial number Status Current status of the instrument Data storage Indicates location of database set where instrument is currently s...

Page 94: ...and Intensity check should be run periodically according to the recommended maintenance schedule See Instrument Diagnostics on page 135 for information about how to run the available instrument diagn...

Page 95: ...select sample and update spectrum select more rows to overlay spectra End experiment and export data Sample names click to edit UV absorbance spectrum for selected samples Measure blank solution Auto...

Page 96: ...limit lines or enter values into the fields at the top of the spectrum Found peaks for the defined range are listed in the table below the spectrum Display options Right click graph to view Display mu...

Page 97: ...icking the sample name field or by importing Sample ID file Edit sample base name After you measure a blank and before the first sample is measured click Sample Name field enter new base name press En...

Page 98: ...the spectrum s vertical axis Micro volume absorbance measurements are normalized to a 10 0 mm pathlength equivalent Measurement Alerts The Acclaro Sample Intelligence technology built into the NanoDr...

Page 99: ...ume Sample and Absorbance Detection Limits Auto Measure and Auto Blank Options Speed up sample analysis with the NanoDrop Eight Auto Measure and Auto Blank features which cause the instrument to start...

Page 100: ...select End Experiment the End Experiment dialog box is displayed Click Save and you will have the option to Print or Export the experiment data Available options Note The End Experiment button is ena...

Page 101: ...PC any USB port or to a network location Available export file formats comma separated values spreadsheet csv file tab separated values spreadsheet tsv file NanoDrop Eight software n8db file The file...

Page 102: ...dot menu above the plate matrix user interface in the bottom left of the measurement acquisition screen 96 well plate allows users to keep track of sample measurements for samples stored in a standar...

Page 103: ...ll either be filled bright green indicating no sample alerts or with an A in a circle indicating an Acclaro alert is present for that sample NanoDrop Eight tube rack accessory Sample channels Measurem...

Page 104: ...ey wish to measure To activate inactive all channels at once click the circle in the column header area next to Sample Alternatively users can activate inactivate individual channels by clicking the c...

Page 105: ...rement data from History Click the menu in the upper right of the measurement screen to access available menu options Home Return to NanoDrop Eight Home screen Import Import data from a USB flash driv...

Page 106: ...ters include time range application type and any user defined labels see Manage Identifiers for information about adding and deleting labels Here is an example Export Selected Experiments Use Select i...

Page 107: ...cess message select OK Delete Selected Experiments Select experiments to be deleted Delete selected experiments Filter experiments as needed use Search feature to find desired experiment Click the che...

Page 108: ...r you open an experiment the software shows the UV or UV visible absorbance spectrum and associated data for the selected experiment much like it appears during a measurement The image below describes...

Page 109: ...add one or more tags i e labels or metadata tags to an experiment to make the experiment easier to find Use the History to add labels to experiments assign existing labels view assigned labels and rem...

Page 110: ...Manage Tags box enter a tag as an Identifier Click Save to add the tag Add additional tags as needed Click OK Remove a label from Home screen select History Click the menu of the experiment row and se...

Page 111: ...automatically Export Selected Experiments You can export measurement data when you save the experiment or afterwards from the History Measurement data can be exported in three formats as comma separat...

Page 112: ...for export each exported experiment has a corresponding file The filenames are the same as the experiment names The files are stored in a folder named NanodropEight followed by the instrument serial...

Page 113: ...export path click Export Export data from History From Home screen open History Click the checkbox for the experiment s you want to export You can use the Search feature to find experiment Click Expo...

Page 114: ...rom any experiment or all the measurements in the database Delete data from any measurement screen click a sample measurement row click Delete data from History from Home screen open History select ro...

Page 115: ...elect Print In Print Review select the data you want to include in the data tables choose Print to confirm in the Print Preview window make sure the correct printer is selected and set other print opt...

Page 116: ...rm in the Print Preview window make sure the correct printer is selected and set other print options as desired such as paper size and orientation Auto setting is recommended margin and alignment to a...

Page 117: ...minant in dsDNA or dsDNA in RNA These models are specific to the source of the nucleic acid The mouse icon is representative of all mammalian sources of nucleic acid If you are measuring nucleic acid...

Page 118: ...pear in the Sample Plate Map the data table and in History see below The icons are active in all three places the information remains with the data indefinitely even after it has been exported Contami...

Page 119: ...e of guanidine HCl and guanidinium isothiocyanate detects species specific dsDNA contamination in the RNA application and detects species specific RNA contamination is in the dsDNA application for pro...

Page 120: ...bance due to phenol 2 53 at the analysis wavelength 260 nm coefficient of variation for the measurement result uncertainty x 100 measurement result 4 22 a high CV indicates the measurement result is c...

Page 121: ...titative approach to contaminant analysis Through sophisticated mathematical algorithms Acclaro analyzes the spectral data to identify probable contaminants in a sample and removes any contribution du...

Page 122: ...haracteristics include absorbance ratios which indicate the presence of compounds that may interfere with sample measurements also referred to as purity ratios For more information watch the multimedi...

Page 123: ...language requires a software restart Database The NanoDrop Eight uses a built in database Security Used when the optional Thermo Scientific SciVault software is installed The server location is autom...

Page 124: ...tein editor Measurement Screen Display Options When performing measurements using the PC Control software right click the graph on the measurement screen to bring up the following display options Disp...

Page 125: ...specified range You can enter the range by dragging the color coded limit lines or enter values into the fields at the top of the spectrum Found peaks for the defined range are listed in the table bel...

Page 126: ...7 Learning Center Measurement Screen Display Options 126 NanoDrop Eight User Guide Thermo Scientific...

Page 127: ...Thermo Scientific NanoDrop Eight User Guide 127 8 Maintenance Maintenance Schedule 128 Maintaining the Pedestals 129 Decontaminating the Instrument 133 Instrument Diagnostics 135...

Page 128: ...ition pedestals Every 6 Months Recondition pedestals Run Intensity Check Run Performance Verification Run Pedestal Image Check If you are experiencing an issue with your system refer to the troublesho...

Page 129: ...cid HF on the pedestals Fluoride ions will permanently damage the quartz fiber optic cables To prevent damage from spills keep containers of liquids away from the instrument Do not use a squirt or spr...

Page 130: ...h a new laboratory wipe To clean the pedestals between users 1 Lift the arm and clean all pedestals with a new laboratory wipe 2 Pipette 3 5 L DI H2O onto the lower pedestal 3 Lower the arm and wait 2...

Page 131: ...ower pedestal to flatten out preventing proper formation of the liquid column when the arm is lowered The resulting spectrum may look rough or jagged If samples flatten out on the pedestal rather than...

Page 132: ...und to dry 3 Fold a clean laboratory wipe into quarters and use it to vigorously buff the surface of each pedestal Notice Support the instrument arm with one hand while you buff the upper pedestal to...

Page 133: ...th one hand while you buff the upper pedestal to avoid damaging the arm 7 Fold a clean laboratory wipe into quarters and use it to vigorously buff the surface of each pedestal at least 50 times 8 Use...

Page 134: ...l with a new laboratory wipe 2 Pipette 2 3 L diluted bleach solution see Supplies needed onto the lower pedestal 3 Lower the arm and wait 2 3 minutes 4 Lift the arm and clean all pedestals with a new...

Page 135: ...iagnostics Thermo Scientific NanoDrop Eight User Guide 135 Instrument Diagnostics Every 6 months run the following performance and quality checks to verify instrument operation Intensity Check 136 Per...

Page 136: ...cifications The test is automatically performed using the pedestal paths Supplies needed lint free laboratory wipes To run intensity check 1 From the Home screen select Intensity 2 Lift the instrument...

Page 137: ...the intensity graphs of a single channel by clicking a single row from the Channel table or Shift Click to highlight multiple rows see image to the right A green check mark and banner will be displaye...

Page 138: ...ansfer pipettes laboratory gloves Before you begin First make sure the pedestals are properly conditioned To test pedestal conditioning clean the upper and lower pedestal surfaces with a dry lint free...

Page 139: ...gid structure to ensure the tips do not splay or skew when delivering the samples Tip It is suggested that this technique be practiced until consistent delivery to all eight positions is routine and q...

Page 140: ...remove each aliquot and use a fresh aliquot of PV 1 for each measurement b In between each measurement remove PV 1 solution from all 16 pedestals upper and lower using a clean dry laboratory wipe 14...

Page 141: ...ic NanoDrop Eight User Guide 141 3 Click End Experiment when done 4 To review results from a previous Performance Verification check select History from the Home screen and locate the Performance Veri...

Page 142: ...8 Maintenance 142 NanoDrop Eight User Guide Thermo Scientific This page is intentionally blank...

Page 143: ...ntific NanoDrop Eight User Guide 143 9 Safety and Operating Precautions Contents Operating Precautions 144 Safety Information 145 Note Be sure that all persons operating this system read the safety ma...

Page 144: ...ous Materials Do not use hydrofluoric acid HF on the pedestals Fluoride ions will permanently damage the quartz fiber optic cables To prevent damage from spills keep containers of liquids away from th...

Page 145: ...n the instrument itself The symbol indicates that there is additional safety information in the documentation and failure to heed the safety precautions could result in injury WARNING Indicates a haza...

Page 146: ...trical shock in the vicinity Only qualified persons shall perform the related procedures Avoid fire hazard Do not test flammable or explosive samples Read and follow the associated instructions carefu...

Page 147: ...injury If this equipment is used in a manner not specified in the accompanying documentation the protection provided by the equipment may be impaired CAUTION Avoid personal injury Perform only those p...

Page 148: ...s damaged contact us Power Line Conditioning Accessories CAUTION Avoid shock hazard Only a qualified person using the appropriate measuring device shall check the line voltage current and frequency On...

Page 149: ...able than the entire system including accessories typically uses Maximum power consumption and heat dissipation specifications for the spectrometer and accessories are shown below The values are appro...

Page 150: ...ve to small volume measurements when using the pedestal for measurements on any of the NanoDrop instruments If you are measuring samples with high vapor pressures use an instrument with provision for...

Page 151: ...m to HCl alcohol bleach acetone or other solvents as the adhesive securing the seal may be affected If the seal comes loose please contact us Biohazard or Radioactive Materials and Infectious Agents B...

Page 152: ...e contaminated with biohazard or radioactive materials infectious agents or any other materials and or conditions that could constitute a health or injury hazard to employees Contact us if you have qu...

Reviews: