9
Safety and Operating Precautions
Safety Information
Thermo Scientific
NanoDrop Eight User Guide 147
When the System Arrives
When the instrument arrives, check the exterior of the shipping box for signs of
damage. If damage is apparent, contact us or your local distributor for instructions.
• Move the shipping box to the installation location at least 24 hours before
installation.
• Keep the instrument upright at all times.
The warranty will not cover:
• Damage due to improper moving techniques.
• Damage due to removing the sealed plastic bag before the instrument has come
to room temperature.
•
Lifting or Moving the Instrument
To avoid risk of injury, use proper lifting techniques when lifting or moving the
instrument or other system components.
WARNING
Avoid personal injury. If this equipment is used in a manner not
specified in the accompanying documentation, the protection provided by the
equipment may be impaired.
CAUTION
Avoid personal injury. Perform
only
those procedures described in
the documentation. If there are other problems, contact us. Any other service
must be performed by trained personnel.
CAUTION
Avoid shock hazard. Do not remove the cover of the instrument. All
service to the instrument must be performed by trained personnel.
Note
• Inside the shipping box, the instrument is sealed in a plastic bag to keep
the unit dry.
• Allow 24 hours for the instrument to reach room temperature before
opening the bag. If the bag is opened before the instrument reaches
room temperature, moisture could condense on the optical components
and cause permanent damage.
Note
It is important to have all system utilities installed before the instrument
arrives. Utility installations must comply with all local building and safety
codes.
Summary of Contents for NanoDrop Eight
Page 6: ...Contents vi NanoDrop Eight User Guide Thermo Scientific...
Page 24: ......
Page 56: ...5 Protein Applications Measure Protein A280 56 NanoDrop Eight User Guide Thermo Scientific...
Page 76: ...6 Custom Applications Measure Custom 76 NanoDrop Eight User Guide Thermo Scientific...
Page 142: ...8 Maintenance 142 NanoDrop Eight User Guide Thermo Scientific This page is intentionally blank...