Herasafe 2030i / Maxisafe 2030i
50155215 |
41
Start-up |
Chapter 5
USB Interface
The USB port on top of the device is reserved for servicing!
Figure 24
USB Interface and Display
The USB port next to the display window enables the user to save log data and install new
software releases.
Installation Test
Do not operate the device before the installation test has been completed.
The installation inspection of the device must be conducted in accordance with EN
12469 and additionally according to DIN 12980 for Maxisafe 2020. The cabinet may
be operated as a Class II microbiological safety cabinet, in accordance with EN
12469 / 2000, if the device functions or function patterns listed below were checked
and if the test results are within the safety value tolerances specified by the
manufacturer:
Electrical safety test
Inflow speed test
Downflow speed test
Leakage test of HEPA filters
Airflow control test
Installation test according to DIN 12980:2016 for Maxisafe 2030i
A repeat test must also be performed after repairs to the device or after considerable
changes (more than 5 cm) to the location of the device.
The operator must prepare a test report or request a written test report from the
authorized test service.
Safety warranty
The operational safety of the device, particularly the personal and material pro-
tection, are guaranteed only if all safety functions of the device have been test-
ed and approved.
Thermo Electron LED will not warrant the operational safety if the device is op-
erated without performance of the required installation test or if the installation
test and repeat test are not performed by adequately trained and authorized
personnel!
Summary of Contents for HERASAFE 2030i Series
Page 1: ...Valid 30 08 2017 50155204 21st September 2021 Safety cabinets 50155215 Rev B...
Page 6: ...Table of Contents 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 10: ...Figures 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 20: ...Delivery 10 50155215 Herasafe 2030i Maxisafe 2030i Chapter 2...
Page 130: ...Operation 120 50155215 Herasafe 2030i Maxisafe 2030i Chapter 6...
Page 134: ...Preparation 124 50155215 Herasafe 2030i Maxisafe 2030i Chapter 7...
Page 136: ...Shut down 126 50155215 Herasafe 2030i Maxisafe 2030i Chapter 8...
Page 144: ...Maintenance 134 50155215 Herasafe 2030i Maxisafe 2030i Chapter 10...
Page 154: ...Technical Data 144 50155215 Herasafe 2030i Maxisafe 2030i Chapter 12...
Page 156: ...Device Log 146 50155215 Herasafe 2030i Maxisafe 2030i Chapter 13...
Page 157: ...Herasafe 2030i Maxisafe 2030i 50155215 147 Certificate of Decontamination...
Page 158: ...Certificate of Decontamination 148 50155215 Herasafe 2030i Maxisafe 2030i Chapter 14...
Page 173: ...Herasafe 2030i Maxisafe 2030i 50155215 163 Appendix City Time Zones Chapter 15...