2.
Insert a small flat-tip screwdriver into the slot in the upper portion of the power entry module, and
open the door.
3.
Pull out the fuse compartment to inspect the two fuses in the fuse compartment:
4.
Pull out the blown fuse(s) from the fuse compartment and replace with new 10 A Type M 250 V 5 ×
20 mm fuses.
5.
Place the fuse compartment back into the power entry module (there is only one orientation) and
close the door.
6.
Press the door until it locks in place.
7.
Connect the instrument power cord.
Parts list
Table 7 Kits and spares
Part
PN
Kits
Temperature Verification Kit (0.1
‑
mL)
4373997
Temperature Verification Kit (0.2
‑
mL)
4377669
Temperature Verification Kit (0.02
‑
mL)
4385338
Temperature Verification Kit (0.5
‑
mL)
4309924
Spares
Cable, AC Power Supply
4374647
Cam, Left Side
4374274
Cam, Right Side
4374275
Chapter 5
Maintenance
Parts list
5
Veriti
™
Thermal Cycler User Guide
81