4.
In the
Admin Menu
screen, select
Restore Factory Settings
to open the
Restore Factory
Settings
screen:
5.
In the
Restore Factory Settings
screen, select
Restore
. The
Restore Factory Settings
dialog
box is displayed.
6.
In the dialog box, select:
•
Yes
to back up run methods and other settings before restoring the default settings to the
instrument. Insert a USB drive for the backup.
•
No
to restore the default settings, without first performing a back up. Before the restore
operation begins, you are prompted to confirm the restore with a second dialog box.
IMPORTANT!
Selecting
No
erases all run methods, user accounts and other settings from
the instrument.
•
Cancel
to cancel the restore operation.
When the restore process is complete, a message box stating “Factory settings have been
restored.” opens.
7.
Remove the USB drive, if present, then toggle the switch on the back of the instrument to power
off.
8.
Toggle the switch again to power on the instrument and restart with the factory default settings.
Running the cycle performance test
Use this test to verify that the sample block(s) are heating and cooling properly and that the PCR
cycling function operates properly.