![Thermo Spectronic BioMate 3 Service Manual Download Page 71](http://html.mh-extra.com/html/thermo/spectronic-biomate-3/spectronic-biomate-3_service-manual_1104233071.webp)
Optical Alignment
Beam Alignment with Microcell in Sample Compartment
Alignment Procedure:
1. Install single cell holder (335916) into the sample compartment.
2. Use a microcell (331752-000); make sure the cell is perfectly clean.
3. Partially (about ½ height) fill the microcell with deionized water.
4. Gently tap the cell on a padded surface to ensure that there are no bubbles in the light path.
Method
(best if performed at zero order)
1. Use Service Diagnostics to continuously pulse the xenon lamp and to access zero order.
2. Place a piece of clean white paper (like the back of a calling card) up against the sample detector
window so that the clean white side faces the single cell holder.
3. Insert the microcell into the single holder.
4. Observe the intensity of light reflecting off the white paper as adjustments of Mirror C are made. Set
for maximum uniform light image intensity on the white paper.
Final Check:
1. Reassemble the unit.
2. Set the wavelength for 260nm.
3. Use a microcell (331752-000); make sure the cell is perfectly clean.
4. Fill the microcell with deionized water.
5. Gently tap the cell on a padded surface to ensure that there are no bubbles in the light path.
6. Insert the cell in the cell holder and zero the instrument.
7. Remove and re-insert the microcell in exactly the same position several times.
8. Confirm
zero
repeatability (should be /- 0.003).
An additional check is to turn the microcell 180 degrees (reverse direction), reinsert the cell and compare
the forward/reverse direction agreeability. There is no specification on this; however, it should have
similar repeatability.
Perform the above ‘Final Check’ at several wavelengths, especially at 260nm and 280nm.
Note:
The ‘axial’ position of the square cuvette holder also affects the reading/repeatability. It
may be necessary to loosen the screw that secures the cuvette and turn it to ensure that
the light beam passes straight through the cell windows on BOTH sides of the cuvette
without hitting the inner side wall of the cell’s sample chamber.
6-7
Summary of Contents for Spectronic BioMate 3
Page 1: ......
Page 6: ...This page intentionally left blank vi ...
Page 13: ...Optical Diagram Figure 1 1 Optical Diagram Shown are both UV Vis and Vis Optics 1 5 ...
Page 14: ...Specifications This page intentionally left blank 1 6 ...
Page 22: ...Troubleshooting Guide This page intentionally left blank 3 6 ...
Page 36: ...Routine Maintenance This page intentionally left blank 4 8 ...
Page 72: ......
Page 78: ...Drawings This page intentionally left blank 8 2 ...
Page 79: ...Drawings 8 3 Figure 8 1 Spectronic GENESYS 10 Wiring Diagram ...
Page 80: ...wings 8 4 This page intentionally left blank Dra ...
Page 81: ...Drawings 8 5 Figure 8 2 Keypad Schematic ...
Page 82: ...wings 8 6 This page intentionally left blank Dra ...
Page 84: ...wings 8 8 This page intentionally left blank Dra ...
Page 86: ...Drawings This page intentionally left blank 8 10 ...
Page 88: ...Drawings This page intentionally left blank 8 12 ...
Page 90: ...Drawings This page intentionally left blank 8 14 ...
Page 92: ...Drawings This page intentionally left blank 8 16 ...
Page 94: ...Drawings This page intentionally left blank 8 18 ...
Page 96: ...Drawings This page intentionally left blank 8 20 ...
Page 98: ...Drawings This page intentionally left blank 8 22 ...
Page 100: ...Drawings This page intentionally left blank 8 24 ...
Page 102: ...Drawings This page intentionally left blank 8 26 ...
Page 104: ...Drawings This page intentionally left blank 8 28 ...
Page 106: ...Drawings This page intentionally left blank 8 30 ...
Page 108: ...Drawings This page intentionally left blank 8 32 ...
Page 110: ...Drawings This page intentionally left blank 8 34 ...
Page 112: ...Drawings This page intentionally left blank 8 36 ...
Page 114: ...Drawings This page intentionally left blank 8 38 ...
Page 116: ...Drawings This page intentionally left blank 8 40 ...
Page 118: ...Drawings This page intentionally left blank 8 42 ...
Page 120: ...Drawings This page intentionally left blank 8 44 ...
Page 122: ...Drawings This page intentionally left blank 8 46 ...
Page 124: ...Drawings This page intentionally left blank 8 48 ...
Page 126: ...Drawings This page intentionally left blank 8 50 ...
Page 128: ...Drawings This page intentionally left blank 8 52 ...
Page 130: ...Drawings This page intentionally left blank 8 54 ...
Page 132: ...Drawings This page intentionally left blank 8 56 ...
Page 134: ...Drawings This page intentionally left blank 8 58 ...
Page 136: ...Drawings This page intentionally left blank 8 60 ...
Page 138: ...Drawings This page intentionally left blank 8 62 ...
Page 140: ...Drawings This page intentionally left blank 8 64 ...
Page 142: ...Drawings This page intentionally left blank 8 66 ...
Page 144: ...Drawings This page intentionally left blank 8 68 ...
Page 146: ...Drawings This page intentionally left blank 8 70 ...
Page 148: ...Drawings This page intentionally left blank 8 72 ...
Page 150: ...Drawings This page intentionally left blank 8 74 ...
Page 152: ...Drawings This page intentionally left blank 8 76 ...
Page 166: ...Service Diagnostic Software This page intentionally left blank 9 14 ...
Page 170: ...Technical Support Bulletins This page intentionally left blank 10 4 ...