Disassembly and Replacement
Replacement
1.
Replace the turret shaft.
2.
Pass the turret shaft with attached driven gear up through the bushing from the bottom of the unit
until the groove for the E-clip is accessible above the top of the bushing. If the turret motor was not
removed. It may be necessary to rotate the gear slightly to make sure the teeth of the driven gear
mesh with the teeth of the drive gear on the motor shaft.
3.
Insert the E-clip into the shaft groove.
4.
Replace the turret motor.
5.
Pass the turret motor shaft through the hole in the base and push it all the way down till the motor
is flush against the base. DO NOT FORCE THE MOTOR! It may be necessary to rotate the motor
slightly to make sure the teeth of the drive gear on the end of the motor shaft mesh properly with
the teeth of the driven gear.
6.
Insert and tighten the two screws, which secure the motor to the base.
7.
Pass the connector end of the motor wires through the hole in the base making sure the other
wires passing through the hole are not damaged.
8.
If all other work is complete, replace the Main Circuit board (refer to "Main board Connections").
9.
Attach the connector on the end of the motor wires to the Main Circuit board.
10.
Replace the turret Opto Switch.
11.
Insert and tighten the two screws, which secure the turret Opto Switch to the base.
12.
Attach the connector on the end of the switch wires to the Main Circuit board.
Light Source (335923-000 [Vis Only], 335902-146S [UV-Vis Only] and 335908
[Spectronic GENESYS 6 Only])
NOTE:
There are two possible light sources for the Spectronic GENESYS 10
Spectrophotometer: xenon (UV-Visible) or tungsten (Visible only). Each type is unique to
a specific instrument they cannot be interchanged. The Vis lamp is supplied as an
assembly already aligned on its mounting brackets and is replaced as a complete
assembly. The UV Lamp is supplied alone and requires optical alignment after replacing.
5-17
Summary of Contents for Spectronic BioMate 3
Page 1: ......
Page 6: ...This page intentionally left blank vi ...
Page 13: ...Optical Diagram Figure 1 1 Optical Diagram Shown are both UV Vis and Vis Optics 1 5 ...
Page 14: ...Specifications This page intentionally left blank 1 6 ...
Page 22: ...Troubleshooting Guide This page intentionally left blank 3 6 ...
Page 36: ...Routine Maintenance This page intentionally left blank 4 8 ...
Page 72: ......
Page 78: ...Drawings This page intentionally left blank 8 2 ...
Page 79: ...Drawings 8 3 Figure 8 1 Spectronic GENESYS 10 Wiring Diagram ...
Page 80: ...wings 8 4 This page intentionally left blank Dra ...
Page 81: ...Drawings 8 5 Figure 8 2 Keypad Schematic ...
Page 82: ...wings 8 6 This page intentionally left blank Dra ...
Page 84: ...wings 8 8 This page intentionally left blank Dra ...
Page 86: ...Drawings This page intentionally left blank 8 10 ...
Page 88: ...Drawings This page intentionally left blank 8 12 ...
Page 90: ...Drawings This page intentionally left blank 8 14 ...
Page 92: ...Drawings This page intentionally left blank 8 16 ...
Page 94: ...Drawings This page intentionally left blank 8 18 ...
Page 96: ...Drawings This page intentionally left blank 8 20 ...
Page 98: ...Drawings This page intentionally left blank 8 22 ...
Page 100: ...Drawings This page intentionally left blank 8 24 ...
Page 102: ...Drawings This page intentionally left blank 8 26 ...
Page 104: ...Drawings This page intentionally left blank 8 28 ...
Page 106: ...Drawings This page intentionally left blank 8 30 ...
Page 108: ...Drawings This page intentionally left blank 8 32 ...
Page 110: ...Drawings This page intentionally left blank 8 34 ...
Page 112: ...Drawings This page intentionally left blank 8 36 ...
Page 114: ...Drawings This page intentionally left blank 8 38 ...
Page 116: ...Drawings This page intentionally left blank 8 40 ...
Page 118: ...Drawings This page intentionally left blank 8 42 ...
Page 120: ...Drawings This page intentionally left blank 8 44 ...
Page 122: ...Drawings This page intentionally left blank 8 46 ...
Page 124: ...Drawings This page intentionally left blank 8 48 ...
Page 126: ...Drawings This page intentionally left blank 8 50 ...
Page 128: ...Drawings This page intentionally left blank 8 52 ...
Page 130: ...Drawings This page intentionally left blank 8 54 ...
Page 132: ...Drawings This page intentionally left blank 8 56 ...
Page 134: ...Drawings This page intentionally left blank 8 58 ...
Page 136: ...Drawings This page intentionally left blank 8 60 ...
Page 138: ...Drawings This page intentionally left blank 8 62 ...
Page 140: ...Drawings This page intentionally left blank 8 64 ...
Page 142: ...Drawings This page intentionally left blank 8 66 ...
Page 144: ...Drawings This page intentionally left blank 8 68 ...
Page 146: ...Drawings This page intentionally left blank 8 70 ...
Page 148: ...Drawings This page intentionally left blank 8 72 ...
Page 150: ...Drawings This page intentionally left blank 8 74 ...
Page 152: ...Drawings This page intentionally left blank 8 76 ...
Page 166: ...Service Diagnostic Software This page intentionally left blank 9 14 ...
Page 170: ...Technical Support Bulletins This page intentionally left blank 10 4 ...