![Thermo Spectronic BioMate 3 Service Manual Download Page 177](http://html.mh-extra.com/html/thermo/spectronic-biomate-3/spectronic-biomate-3_service-manual_1104233177.webp)
Technical Support Bulletins
S-399 - Spectronic GENESYS 10 and Spectronic BioMate 3: New Firmware
October 2001
REFERENCE
: ECO
71863
PROBLEM:
1. Earlier revision of firmware will not work with new EDT display (335901-
657S).
2. Error in Performance Validation test – Internal Wavelength tolerance (is
±
1nm at 883nm, should be
±
2nm).
SYMPTOM/
CAUSE:
Problem 1 - can cause the display to be blank when the unit is turned on (usually
after powering-down an instrument that has been on beyond the warm-up period).
Problem 2 - can cause the instrument to fail while still within the published
specifications.
SOLUTION:
Release of new firmware: 1.103 for Spectronic GENESYS 10; 1.001 for
Spectronic BioMate 3
PRODUCT
AFFECTED:
All Spectronic GENESYS 10 spectrophotometers of Serial Number < ###D260###
were manufactured with firmware <1.103 rev.
All Spectronic BioMate 3 spectrophotometers of Serial Number < ###D260### were
manufactured with firmware <1.001 rev.
APPLIES TO:
Repair at customer site
Repair at Depot
The new firmware will work with the new EDT display (see TSB S-391), and has the
correct Internal Wavelength Tolerance of
±
2nm at 883nm.
If the unit has the new EDT display which shows the problem above, the unit should
be sent to the Service depot for upgrading since the upgrade will require the
installation of the ROM IC and wavelength recalibration using the appropriate
standards and software.
11-7
Summary of Contents for Spectronic BioMate 3
Page 1: ......
Page 6: ...This page intentionally left blank vi ...
Page 13: ...Optical Diagram Figure 1 1 Optical Diagram Shown are both UV Vis and Vis Optics 1 5 ...
Page 14: ...Specifications This page intentionally left blank 1 6 ...
Page 22: ...Troubleshooting Guide This page intentionally left blank 3 6 ...
Page 36: ...Routine Maintenance This page intentionally left blank 4 8 ...
Page 72: ......
Page 78: ...Drawings This page intentionally left blank 8 2 ...
Page 79: ...Drawings 8 3 Figure 8 1 Spectronic GENESYS 10 Wiring Diagram ...
Page 80: ...wings 8 4 This page intentionally left blank Dra ...
Page 81: ...Drawings 8 5 Figure 8 2 Keypad Schematic ...
Page 82: ...wings 8 6 This page intentionally left blank Dra ...
Page 84: ...wings 8 8 This page intentionally left blank Dra ...
Page 86: ...Drawings This page intentionally left blank 8 10 ...
Page 88: ...Drawings This page intentionally left blank 8 12 ...
Page 90: ...Drawings This page intentionally left blank 8 14 ...
Page 92: ...Drawings This page intentionally left blank 8 16 ...
Page 94: ...Drawings This page intentionally left blank 8 18 ...
Page 96: ...Drawings This page intentionally left blank 8 20 ...
Page 98: ...Drawings This page intentionally left blank 8 22 ...
Page 100: ...Drawings This page intentionally left blank 8 24 ...
Page 102: ...Drawings This page intentionally left blank 8 26 ...
Page 104: ...Drawings This page intentionally left blank 8 28 ...
Page 106: ...Drawings This page intentionally left blank 8 30 ...
Page 108: ...Drawings This page intentionally left blank 8 32 ...
Page 110: ...Drawings This page intentionally left blank 8 34 ...
Page 112: ...Drawings This page intentionally left blank 8 36 ...
Page 114: ...Drawings This page intentionally left blank 8 38 ...
Page 116: ...Drawings This page intentionally left blank 8 40 ...
Page 118: ...Drawings This page intentionally left blank 8 42 ...
Page 120: ...Drawings This page intentionally left blank 8 44 ...
Page 122: ...Drawings This page intentionally left blank 8 46 ...
Page 124: ...Drawings This page intentionally left blank 8 48 ...
Page 126: ...Drawings This page intentionally left blank 8 50 ...
Page 128: ...Drawings This page intentionally left blank 8 52 ...
Page 130: ...Drawings This page intentionally left blank 8 54 ...
Page 132: ...Drawings This page intentionally left blank 8 56 ...
Page 134: ...Drawings This page intentionally left blank 8 58 ...
Page 136: ...Drawings This page intentionally left blank 8 60 ...
Page 138: ...Drawings This page intentionally left blank 8 62 ...
Page 140: ...Drawings This page intentionally left blank 8 64 ...
Page 142: ...Drawings This page intentionally left blank 8 66 ...
Page 144: ...Drawings This page intentionally left blank 8 68 ...
Page 146: ...Drawings This page intentionally left blank 8 70 ...
Page 148: ...Drawings This page intentionally left blank 8 72 ...
Page 150: ...Drawings This page intentionally left blank 8 74 ...
Page 152: ...Drawings This page intentionally left blank 8 76 ...
Page 166: ...Service Diagnostic Software This page intentionally left blank 9 14 ...
Page 170: ...Technical Support Bulletins This page intentionally left blank 10 4 ...