Measure dsDNA, ssDNA or RNA
18
NanoDrop One User Guide
Thermo Scientific
Related Topics
•
Calculations for Nucleic Acid Measurements
Setting
Available Options
Description
Baseline Correction
On or off
Enter baseline correction
wavelength in nm or use
default value (340 nm)
Optional user-defined baseline correction
. Can be used to
correct for any offset caused by light scattering particulates by
subtracting measured absorbance at specified baseline correction
wavelength from absorbance values at all wavelengths in sample
spectrum. As a result, absorbance of sample spectrum is zero at
specified baseline correction wavelength.
The nucleic acid applications use the
to correlate absorbance with concentration.
Solving Beer’s law for concentration yields the equation
at the right.
Beer-Lambert Equation (solved for concentration)
c = A / (
* b)
where:
A
= UV absorbance in absorbance units (AU)
= wavelength-dependent molar absorptivity coefficient (or extinction
coefficient) in liter/mol-cm
b
= pathlength in cm
c
= analyte concentration in moles/liter or molarity (M)
Note
: Dividing the measured absorbance of a sample solution by its molar
extinction coefficient yields the molar concentration of the sample. See
Published Extinction Coefficients
for more information regarding molar
vs. mass concentration values.
Summary of Contents for NanoDrop One
Page 18: ......
Page 28: ......
Page 58: ......
Page 134: ...Measure Protein Pierce 660 128 NanoDrop One User Guide Thermo Scientific Instrument Settings ...
Page 144: ......
Page 152: ......
Page 159: ...Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Page 160: ...Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Page 239: ...3 Learning Center Instrument Settings Thermo Scientific NanoDrop One User Guide 239 ...
Page 312: ...7 Contact Technical Support 314 NanoDrop One User Guide Thermo Scientific ...