3-6 Thermo
Electron
January
2004
Select Properties from the Method pull-down menu and click on the
Options tab. Uncheck the ‘Analyze during acquisition’ item if it has a
check in the check box (Figure 3.4). Place a check in the check box
for ‘Analyze after acquisition’.
Figure 3.4 Method Properties – Options dialog box
Close the Method Properties Window and click the save button on the
Command Toolbar. Select Save Method and give the method a name
if required. Close the Method Custom Report Window. Verify that
the default printer is properly selected by clicking on File-Print from
the Menu bar. If the correct printer is selected, click on the Cancel
button.
PREVIEW
Verify that the system is fully equilibrated by monitoring the baseline
before the sample is injected. To monitor the baseline click on the
Preview button on the Command Toolbar at the top of the
Applications Window. The baseline is displayed in real-time for all
signals configured in Method-Instrument Setup. Each discrete
wavelength will have its own window. The scan surface will be
displayed in the Mixed View screen as a contour plot, a real-time
spectrum display, a wavelength-selectable chromatogram and the
peak purity index window. The Mixed View screen is easily adjusted
to emphasize any of the four displays.
Summary of Contents for Finnigan SpectraSYSTEM UV6000LP
Page 1: ...Finnigan SpectraSYSTEM UV6000LP Detector Reference Manual A0099 595 Revision F ...
Page 8: ......
Page 12: ......
Page 16: ......
Page 44: ......
Page 76: ......
Page 91: ...January 2004 Thermo Electron 5 15 Figure 5 8 Holmium oxide absorbance spectrum ...
Page 118: ......
Page 129: ...January 2004 Thermo Electron 8 11 PD Z115E Figure 8 6 Power button and bracket removal ...
Page 134: ......