Servicing
Flow Transducer Replacement
6-34
Model 49
i
Primary Standard Instruction Manual
Thermo Fisher Scientific
9. Wait at least 10 seconds for the ambient reading to stabilize, use
and
to enter the known barometric
pressure, and press
to save the pressure value.
10. Reconnect the instrument tubing to the pressure transducer.
11. Install the cover.
Flow Transducer
Replacement
Use the following procedure to replace the flow transducer.
Equipment Required:
Flow transducer
Philips screwdriver
Wrench, 5/16-inch
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Disconnect plumbing from the flow transducer.
3. Disconnect the flow transducer cable (FLOW A or FLOW B) from the
measurement interface board.
4. Loosen the retaining nuts (2) and remove the faulty flow transducer
from the mounting bracket.
5. Install the new flow transducer by following the previous steps in
reverse.
6. Calibrate the flow transducer. Refer to the “Flow Transducer
Calibration” procedure that follows.
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.