Servicing
I/O Expansion Board (Optional) Replacement
6-18
Model 49
i
Primary Standard Instruction Manual
Thermo Fisher Scientific
Note
If Service is not displayed, refer to “
, then return to the beginning of this step.
4. Wait at least 10 seconds for the reading to stabilize, use
and
to enter the known temperature, and press
to
save the temperature value.
5. Install the cover.
I/O Expansion Board
(Optional)
Replacement
Use the following procedure to replace the optional I/O expansion board
(
Note
After replacing the optional I/O expansion board, calibrate the current
outputs and the analog voltage inputs. See the “Analog Output Calibration”
procedure and the “Analog Input Calibration” procedure in this chapter.
Equipment Required:
I/O expansion board
Nut driver, 3/16-inch
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Unplug the I/O expansion board cable from the EXPANSION BD
connector on the motherboard.
3. Remove the two standoffs holding the I/O expansion board connector
to the rear panel (
4. Pop off the board from the mounting studs and remove the board.
5. To install the I/O expansion board, follow previous steps in reverse.
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.