Servicing
I/O Expansion Board (Optional) Replacement
7-40
Model 42
i
Low Source Instruction Manual
Thermo Electron Corporation
4. Push the replacement scrubber into the mounting clips.
5. Attach the Teflon tubing at both ends of the scrubber.
6. Replace the cover.
Figure 7-19.
Replacing the Scrubber
I/O Expansion Board
(Optional)
Replacement
Use the following procedure to replace the optional I/O expansion board
(
Equipment Required:
I/O expansion board
Nut driver, 3/16-inch
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Unplug the I/O expansion board cable from the EXPANSION BD
connector on the motherboard.
Silica Scrubber
Ammonia Scrubber
Mounting Clips
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.