- 30 -
7.2.3
EXAMPLE PROGRAMMING
To show an example programming, the following sintering program is
entered on program number 5 of the controller.
1.
Heating at 600°C/h to 900°C target temperature
2.
Hold for 0.5 h
3.
Heating at 200°C/h to 1,450°C target temperature
4.
Hold for 2.0 h
5.
Heating at 600°C/h to 900°C target temperature
6.
End of program and natural cooling
Table 09: Example programming
Requirements:
The furnace does not heat up.
Push the PROGRAM button
until the Program / Segment display shows “P5”.
Wait approx. 35 seconds until P5
is replaced by a flashing number.
Push the LEFT ARROW button
until the Program / Segment display shows “00”.
Segment “00” in each program allows
setting a program delay using the format hh.mm
(hour hour.minute minute) to
realize a timer start. Entering the value “0.00” starts the
controller immediately after pushing the button.
Push the RIGHT ARROW button
once to change to segment 01.
Define a ramp
(see 6.2.2 and 6.2.1) using the
respective parameters.
Push the RIGHT ARROW button
once to change to segment 02.
Define a dwell time
(see 6.2.2 and 6.2.1) using the
respective parameters.