5
•
Installation
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 85
5.8
Fraction Collection Valve
There are three capillaries or tubing that are required to assembly to the
fraction collection valve when installing the device for the first time.
•
Flush Solvent tubing (to flush solvent reservoir)
•
Delay capillary
•
Needle capillary (to needle unit inside fraction collector
compartment)
5.8.1
Port assignments of the Fraction Collection Valve
Figure 31: Fraction collection valve with connected capillaries except port
6
The fraction collection valve ports are assigned as follows:
Port
Connected Component
Port
Connected Component
1
to flush buffer loop
(pre-installed)
5
to Y-connector
(pre-installed)
2
to needle Capillary
6
Central port, inlet from detector
3
to Y-connector via air
bubble sensor (pre-installed)
7
to flush solvent
4
to flush buffer loop
(pre-installed)
8
to flush pump
(pre-installed)
Summary of Contents for VF-F20-A
Page 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Page 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Page 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Page 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Page 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...