5.6.3
Connecting Fittings, Capillaries, and Tubing
This section provides information about how to connect and handle
capillaries, fittings, and tubing.
5.6.3.1
General Guidelines
When connecting capillaries and tubing, follow these general
recommendations:
•
Use only the capillaries and tubing (for example, solvent lines or
waste tubing) that are shipped with the pump or additional or spare
capillaries and tubing as recommended by Thermo Fisher Scientific.
•
The connectors must be free from contaminants. Even minute
particles may cause damage to the system or lead to invalid test
results.
•
Do not install capillaries or tubes that are stressed, nicked, kinked, or
otherwise damaged.
•
Install capillaries and fittings only at the positions for which they are
intended.
5.6.3.2
Connecting Viper Capillaries
This section describes how to connect Viper™ capillaries. All Viper flow
connections in the Vanquish system are designed to be finger-tight.
To connect Viper capillaries with knurl, follow these steps:
NOTICE
•
Tighten or loosen Viper capillaries
only
with your fingers. Do not use
tools other than the knurl that comes with the capillary.
•
To avoid damage to the capillary or connection, tighten and loosen
the Viper connections
only
when the system pressure is down to
zero.
Summary of Contents for Vanquish VH-P10-A-02
Page 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...