5
•
Installation
Page 60
Vanquish Pumps (VH-P10-A-02) Operating Manual
1. Connect the pump to the drain system (see
2. Connect the solvent lines (see
3. Set up the seal wash system (see
4. Connect the pump to the autosampler (see
NOTICE
When you install devices or components to the system, always
flush them to waste before connecting them in the system flow path. To
flush the Vanquish modules, follow the instructions in the
Vanquish
System Operating Manual
.
5.6.2
Guiding Capillaries and Tubing Through the System
Flow connections between the modules of the Vanquish system are
guided through either the tubing chase in the devices or the guide holes
or capillary clips of the devices.
Tubing Chase, Tubing Guide, Tubing Bracket
To guide certain tubes and lines (solvent tubing, wash liquid tubing,
detector waste line) from the top module to the bottom module in the
Vanquish system stack, the stackable modules have a tubing chase on
the inside right.
The tubing chase provides four tubing guides. Each guide can hold up to
three tubes or lines.
In each module, push the tube (or line) into the appropriate guide (see
picture).
Summary of Contents for Vanquish VH-P10-A-02
Page 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...