8
•
Troubleshooting
Vanquish Pumps (VH-P10-A-02) Operating Manual
Page 213
8.5.3
Performing Pressure Recalibration
When
Messages regarding the pump head pressure or nonlinear pressure
sensor appear in chromatography software, for example:
•
Code 4212 Nonlinear pressure sensor. Please perform pressure
recalibration.
•
Code 4213 Head pressure does not match system pressure signal.
Execute self-test or perform pressure recalibration.
Follow these steps
1.
Before
you perform pressure recalibration, check the message in the
Messages
section in this manual.
Before
pressure recalibration should be performed, message-specific
remedial actions might be required. Perform these actions as
applicable and proceed with the next step.
2. On the purge unit, remove the capillary from the
OUT
port.
3. Close the port and the open end of the capillary with an appropriate
plug or cap.
4. In Chromeleon, on the
Service
sub-panel for the pump, click
Calibrate
to start the pressure recalibration procedure. A wizard
guides you through the procedure.
Summary of Contents for Vanquish VH-P10-A-02
Page 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...