5
•
Installation
Vanquish Charged Aerosol Detectors Operating Manual
Page 71
Follow these steps
To set up the liquid flow connections and complete the installation of
the detector, follow these steps:
1. Connect the waste line (see
2. Connect the inlet capillary to the nebulizer (see
section 5.6.5,
page 76
3. Connect the detector to the drain system (see
section 5.6.6,
page 80
TIP
Components or connections in the flow path to other system
modules may be closed with plugs to protect the component or
connection during transport. When you remove the plugs to connect the
detector in the system, keep the plugs. You may need them to close the
connections again, for example, for future transport.
5.6.2
Guiding Capillaries and Tubing Through the System
Flow connections between the modules of the Vanquish system are
guided through either the tubing chase in the devices or the guide holes
in the device housings.
Tubing Chase, Tubing Guide, Tubing Bracket
To guide certain tubes and capillaries (solvent tubing, wash liquid tubing,
detector waste capillary) from the top module to the bottom module in
the Vanquish system stack, the stackable modules have a tubing chase
on the inside right.
The tubing chase provides four tubing guides. Each guide can hold up to
three tubes or capillaries.
Summary of Contents for Vanquish VF-D20
Page 4: ...Contacting Us Page 4 Vanquish Charged Aerosol Detectors Operating Manual...
Page 28: ...2 Safety Page 28 Vanquish Charged Aerosol Detectors Operating Manual...
Page 38: ...3 Detector Overview Page 38 Vanquish Charged Aerosol Detectors Operating Manual...
Page 44: ...4 Unpacking and Transport Page 44 Vanquish Charged Aerosol Detectors Operating Manual...
Page 150: ...7 Maintenance and Service Page 150 Vanquish Charged Aerosol Detectors Operating Manual...
Page 177: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 177 12 Index...
Page 181: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 181...