7
•
Maintenance and Service
Page 120
Vanquish Charged Aerosol Detectors Operating Manual
7.4
Routine and Preventive Maintenance
Optimum detector performance, maximum uptime of the detector, and
accurate results can be obtained only if the detector is in good condition
and properly maintained.
7.4.1
Maintenance Plan
Perform the maintenance procedures in the table on a regular basis. The
frequency given in the table is a suggestion. The optimum frequency for
maintenance depends on several factors, such as the types and amounts
of samples and solvents used with the detector.
Frequency
What you should do...
Daily
Inspect the flow connections for signs of leakage or
blockage.
When using buffer solutions, flush the detector thoroughly
after use with an appropriate solvent that does not
contain buffers or salts.
The stability of the mobile phase may decrease over time.
Prepare mobile phases freshly on a daily basis.
Regularly
Inspect the flow connections for damage, such as cracks,
nicks, cuts, or blockage.
Check that all warning labels are still present on the
detector and clearly legible. If they are not, contact
Thermo Fisher Scientific for replacement.
Annually
Have Thermo Fisher Scientific service personnel perform
preventive maintenance once a year.
TIP
Chromeleon supports functions for estimating the lifetime of
consumables (see
section 7.4.4 Predictive Performance, page 123
).
Summary of Contents for Vanquish VF-D20
Page 4: ...Contacting Us Page 4 Vanquish Charged Aerosol Detectors Operating Manual...
Page 28: ...2 Safety Page 28 Vanquish Charged Aerosol Detectors Operating Manual...
Page 38: ...3 Detector Overview Page 38 Vanquish Charged Aerosol Detectors Operating Manual...
Page 44: ...4 Unpacking and Transport Page 44 Vanquish Charged Aerosol Detectors Operating Manual...
Page 150: ...7 Maintenance and Service Page 150 Vanquish Charged Aerosol Detectors Operating Manual...
Page 177: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 177 12 Index...
Page 181: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 181...