Universal Chromatography Interfaces
UCI-50 and UCI-100
Page
50
Operating Instructions
8.3 Digital Inputs
Fig. 19: Digital inputs (9-pin D-Sub female connector)
Pin
Port
Signal Level
1
Input-1
TTL
2
Input-2
TTL
3
Input-3
TTL
4
Input-4
TTL
5
Input-5
TTL
6
Input-6
TTL
7
Input-7
TTL
8
Input-8
TTL
9
Ground
GND
Tip:
Internal device resistors keep the digital input level at + 5 V (pull-up
resistance). Therefore, it is possible to use a contact closure relay, without an
additional voltage source. This closure has to be installed between GND and
the respective digital input. Once the relay is shut, the digital voltage is
grounded.
The relay is under a low current of about 50 µA. If the signal source is active,
that is, emitting voltage, verify the correct polarization of the digital input. The
ground of the signal source must be connected to the ground of the UCI. The
digital output of the signal source must be connected to the digital input of the
UCI.
Summary of Contents for UCI-100
Page 2: ...Universal Chromatography Interfaces UCI 50 and UCI 100 Operating Instructions...
Page 4: ...Universal Chromatography Interfaces UCI 50 and UCI 100 Page II Operating Instructions...
Page 10: ...Universal Chromatography Interfaces UCI 50 and UCI 100 Page 4 Operating Instructions...
Page 48: ...Universal Chromatography Interfaces UCI 50 and UCI 100 Page 42 Operating Instructions...
Page 52: ...Universal Chromatography Interfaces UCI 50 and UCI 100 Page 46 Operating Instructions...
Page 54: ...Universal Chromatography Interfaces UCI 50 and UCI 100 Page 48 Operating Instructions...
Page 60: ...Universal Chromatography Interfaces UCI 50 and UCI 100 Page 54 Operating Instructions...