Universal Chromatography Interfaces
UCI-50 and UCI-100
Page
14
Operating Instructions
5.
Click
Close
to close the Server Monitor window. The Server Monitor icon
appears on the taskbar.
(Clicking the
Quit Monitor
button quits (exits) the
Server Monitor
program, but does
not stop the server. To stop the server, click
Stop
.)
6.
Use the power unit and power cord shipped with the UCI to connect the instrument to
a grounded power source.
7.
Windows Vista, Windows 7, and Windows
Server 2008
will automatically detect the new UCI and perform the USB installation.
If Windows fails to detect the UCI and launches a wizard instead, this indicates that you
connected the UCI to the computer and turned on the power for the first time
before
you
installed Chromeleon. To resolve the problem:
a)
Click Cancel to exit the wizard.
b)
Disconnect the UCI from the power source.
c)
Install Chromeleon.
d)
Reconnect the UCI to the power source. Windows will now detect the UCI and install
the USB software for the UCI automatically.
Windows XP
will automatically detect the new UCI and launch the Found New Hardware Wizard,
which guides you through the USB installation. Select the following options:
a)
If asked whether Windows can connect to Windows Update to search for software,
select
No, not this time
.
b)
Accept the default option (
Install the software automatically
) and click
Next>
.
c)
Click
Finish
when the wizard reports that the software for the UCI has been
installed.
If Windows fails to detect the UCI and a message box asks for a USB configuration file
(cmwdmusb.inf), this indicates that you connected the UCI to the computer and turned on
the power for the first time
before
you installed Chromeleon. To resolve the problem:
a)
Click
Cancel
in the Windows message box.
b)
Disconnect the UCI from the power source.
c)
Install Chromeleon.
d)
Reconnect the UCI to the power source. Windows will now automatically detect the
UCI and launch the
Found New Hardware Wizard
.
8.
Set up the UCI in Chromeleon (
→
page 27).
Summary of Contents for UCI-100
Page 2: ...Universal Chromatography Interfaces UCI 50 and UCI 100 Operating Instructions...
Page 4: ...Universal Chromatography Interfaces UCI 50 and UCI 100 Page II Operating Instructions...
Page 10: ...Universal Chromatography Interfaces UCI 50 and UCI 100 Page 4 Operating Instructions...
Page 48: ...Universal Chromatography Interfaces UCI 50 and UCI 100 Page 42 Operating Instructions...
Page 52: ...Universal Chromatography Interfaces UCI 50 and UCI 100 Page 46 Operating Instructions...
Page 54: ...Universal Chromatography Interfaces UCI 50 and UCI 100 Page 48 Operating Instructions...
Page 60: ...Universal Chromatography Interfaces UCI 50 and UCI 100 Page 54 Operating Instructions...