Thermo Scientific TZ-32 Zonal Rotor
Cleaning of Rotor and Septa — 44
6. 4.
Handling of Rulon Seal and Renewal
Rulon seal is a soft plastic with excellent lubrication developed by mixing tetrafluoroethylene resin with a special filler, and
is easily scratched by a fingernail. Do not scratch the upper surface, or solution may leak due to poor sealing.
•
Renewal of Rulon seal
Slightly scratched or worn Rulon seal can be reconditioned by polishing. The flat portion of the zonal rotor cover can
used for polishing the Rulon seal. Some technique is required for finishing.
Put fine sandpaper (No. 800 is recommendable) on a rotor cover, and polish the contact surface until it becomes
even.
Do this carefully while removing polishing agent from the Rulon seal surface and sandpaper. Finish it after making
sure no scratch due to the sandpaper remains on the Rulon seal surface.
•
Removal of Rulon seal
The Rulon seal can be washed and sterilized together with the shaft. Remove the Rulon seal if the Rulon seal is
scratched. Be careful not to scratch the Rulon seal surface.
Removing procedure
a. Insert the accessory extractor into two holes.
b. Turn it clockwise and counterclockwise while lightly depressing it.
c. Carefully pull it upwards while turning.
Before resetting it, apply a thin coat of vacuum grease to the O-ring and insert it carefully with soft portions
of applied fingers.
Figure 41
Polishing the Rulon
seal
Figure 42
Removal of the
Rulon seal